9
•
For sterilization, use Steris® or STERRAD® low temperature
sterilization by processing. Temperatures should not exceed
140
°
F (53
°
C) and/or 28 minute cycles. Refer to Table 1.
NOTE:
•
Given the electrical componentry, the video blade should not be
exposed to extreme temperatures during cleaning.
•
DO NOT AUTOCLAVE THE VIDEO HANDLE, BLADE OR WAND.
•
Do not use abrasive cleaning techniques or ultrasonically clean.
•
Ensure the handle, blade and wand are dry & fully checked prior
to storage
Handles, blades and wands should always be tested after cleaning
and prior to use.
Disposable Sheath
The sheath is disposable and intended for single patient use. Do not
reuse or sterilize. Dispose of after each use.
Video Accessories
All cables and other ancillary devices should be cleaned by wiping with a
70% isopropyl alcohol solution, or a mild detergent and water.
Do not immerse these items in fluids. Care should be taken to prevent
excessive moisture from penetrating transmitter and display housings
VLS 8800 Monitor
The display is not waterproof. The face of the unit, including the front
control buttons, can be cleaned with anti-microbial solution wipes.
Cleaning of the unit should be done carefully to avoid introducing
liquids into the interior of the monitor.