background image

 

•  In live video mode, Press “Menu” button 

•  Press Right or Left Hand arrow button to highlight 

“Time” 

•  Press “Enter” button 

•  Press Right or Left arrow buttons to advance to desired 

portion of date and time to set 

•  Press Up or Down arrow buttons to adjust values 

•  Press “Enter” button to return to prior menu 

Photo and Movie quality can be adjusted between standard 

and high quality images using similar actions in this menu. 

 

Press “Menu” to exit to live video 

IV. Cleaning Instructions 
Video Handle, Blades and Wands 

Reusable laryngoscope video blades must undergo high level 

disinfection or sterilization between uses. Laryngoscope video handles 

and wands should be cleaned with low level disinfectant between 

patients; high level disinfection is optional. Always follow hospital 

protocol for cleaning reusable laryngoscope components.  

 

Detach the video blade or wand from the laryngoscope handle. 

If applicable, detach video display from handle prior to cleaning. 

 

Remove batteries from the handle and replace the cap of 

handle.  

 

Clean external surfaces of the blade, wand and handle with a 

damp cloth or soft brush using warm water and soap, isopropyl 

alcohol solution or other anti-microbial disinfectant to remove 

any deposits.  A pH neutral cleaning solutions recommended.  

 

For high-level disinfection use Cidex® 2.4% glutaraldehyde 

solution or similar disinfectant. Follow manufacturer’s usage 

instructions. To avoid damage, do not soak handles in 

disinfectant or washing solutions for 

20 minutes. 

Содержание IntuBrite VLS 8800

Страница 1: ...1 IntuBrite VLS 8800 Video Laryngoscope Operating Instructions Salter Labs 30 Spur Drive El Paso TX 79906 USA www salterlabs com 800 421 0024 customercare salterlabs com L052 0912 Rev D...

Страница 2: ...bly and Disassembly 3 II Operating the VLS 8800 Monitor 4 III VLS 8800 System Operation 5 IV Cleaning Instructions 8 V Batteries and Battery Maintenance 11 VI Troubleshooting 13 VII Warranty Informati...

Страница 3: ...remove unhook from locking tabs and slide off 3 Orient the connector of the VLS 8800 video cable to the video connection on the top of the VLS handle Position the red marks on the connector to align...

Страница 4: ...r would be verified by the illuminated display of the monitor This setup also charges the battery The unit will automatically switch to battery power when power cord is disconnected Startup 1 Assemble...

Страница 5: ...images recorded on the SD card when the unit is in Review mode The right hand Red Record button starts and stops video recording A Recording saving and viewing images and video Still images and video...

Страница 6: ...can be played on the display through communication with the attachable video DVR remote control See Directions for Remote Control DVR Functions in next section B Managing SD card files with the Remote...

Страница 7: ...iles Press Access button Press Right or Left arrow button to advance through images files To Play a selected video file Press Down Arrow button to start Press Down Arrow button again to pause stop To...

Страница 8: ...ands should be cleaned with low level disinfectant between patients high level disinfection is optional Always follow hospital protocol for cleaning reusable laryngoscope components Detach the video b...

Страница 9: ...d after cleaning and prior to use Disposable Sheath The sheath is disposable and intended for single patient use Do not reuse or sterilize Dispose of after each use Video Accessories All cables and ot...

Страница 10: ...e based on procedures compatible with specific materials Disinfection must be performed to approved hospital protocol Salter Labs cannot warrant that any of the recommended methods will guarantee ster...

Страница 11: ...rage In addition the monitor displays a battery level symbol in the lower left hand corner 3 bars fully charged 2 bars 1 2 charged A blinking symbol indicates unit will shut down in several seconds Th...

Страница 12: ...the cap on the top of the handle When connected to a monitor via the cable the VLS handle will operate the VLS blade or wand without a 3 7 volt AA size lithium battery installed However for longer ove...

Страница 13: ...battery switch on the left side of the case is in the ON position and the LED illuminates Green 2 Monitor is powered but shows blue screen Check connection of video blade or wand to handle Ensure blad...

Страница 14: ...tive will need the part number s and serial number s of the items for which repair is being requested under the warranty The Customer Care Representative will review the situation and if appropriate i...

Страница 15: ...the original VLS system an extended warranty package can be purchased separately and independently for each item under the above terms If a VLS system fails for any reason during the extended warranty...

Отзывы: