Page C-1
3.
If the screen does not show “10 0”, press the TARE key once. The indicator is
prompting you to place ten identical items on the platform.
NOTE:
If you wish to change the sample number, simply press the NET/GROSS key
repeatedly until the desired sample number appears. Available choices are 5, 10, 25,
50 and 100.
4.
Place the sample items on the platform all at once and allow the weight indication to
stabilize. Once this is done, the zero indicated after the sample number will change to
a “–“. For example, “10 –“.
5.
Press the NET/GROSS key to take the sample. If the sample size is large enough, the
indicator now displays the number of pieces on the platform. If it does not, the
indicator briefly displays “Add” and automatically increments the sample size. Repeat
Step #4 with the new sample size.
NOTE:
If the indicator continues to display “Add” even after sampling 100 pieces, the
unit weight of the items you wish to count is too light for your scale to process
accurately.
6.
To exit the piece count mode, press the UNITS key. The APW will remain in scale
memory until you repeat Steps # 2 or #3.