background image

20

APPENDIX D: NOTES ON RECHARAGEABLE BATTERY

D.1 OVERVIEW

Your scale contains two internal lead-acid rechargeable batteries connected in parallel.

Before using the scale for the first time, please charge the battery overnight.
The scale’s battery should operate for about 40 hours if left on continuously. Therefore,

greater usage times can be achieved by selecting an appropriate Auto Power Off Period

under 

A15

 of the User Menu.

The battery can be charged while ON or OFF and the scale can be operated while it’s

charging.

D.2 WHEN TO CHARGE THE INTERNAL BATTERY

1. 

The best time to charge the sealed lead-acid type battery is any time the scale is not in

use. You do not have to wait for the Low Battery Indication – in fact it’s best you don’t.

Charging the battery when the scale is not in use keeps the battery “fresh” and is the

recommended way to manage your scale’s battery.

2. 

When the battery needs to be charged, the Low Battery Indicator will appear in the

upper left-hand corner of the display. The scale may be used for an additional 15 to 30

minutes without damage to the internal battery. 

If the scale is used for more than 30

minutes after a low battery indication, permanent damage may occur to the

battery and/or the battery may not accept a charge.

NOTE:

 When a low battery indication occurs, the scale automatically shuts off after 5

minutes of idle use.

D.3 HOW TO CHARGE THE INTERNAL BATTERY

1.  Connect  the  charger  (AC  Adapter  12  VDC,  800mA)  to  the  scale,  and  then  plug  the

charger  into  an  AC  outlet. 

Make  sure  that  the  AC  voltage  appearing  at  the  wall

outlet matches the input voltage marked on the AC adapter.

2.  After the charging period expires, unplug the charger from the AC outlet, then from the

scale. The scale is now ready for use under its own battery power.

NOTE:

 The charger may be left connected to the scale indefinitely without damage to

the internal battery.

D.4 HOW LONG TO CHARGE THE INTERNAL BATTERY

In general, the battery should be allowed to charge a minimum of 1.5 hours for every hour

of use.
If you discharge the battery below 50% and do not allow the proper time for charging, you

may notice a decline in the usage period. This is due to the battery’s reluctance to accept a

charge.

Содержание PALLET RUNNER 3000

Страница 1: ...ruary 13 2004 Contents subject to change without notice Salter Brecknell Weighing Products 1000 Armstrong Drive Fairmont MN 56031 Tel 800 637 0529 Tel 507 238 8702 Fax 507 238 8271 E mail sales salter...

Страница 2: ...n Item 7 3 3 2 Taring an Item 7 8 Chapter 4 Advanced Features and Operation 9 4 1 Piece Counting 9 Chapter 5 Scale Configuration 10 5 1 Scale Configuration Overview 10 11 5 2 User A Menu 12 5 2 1 Ente...

Страница 3: ...internal battery 20 D 4 How long to charge the internal battery 20 LIST OF FIGURES 1 1 PALLET RUNNER 3000 Front Panel 4 3 1 PALLET RUNNER 3000 Display Detail 6 3 2 Function Keys Layout 7 5 1 PALLET RU...

Страница 4: ...ed after unpacking the scale A small amount of preparation is necessary before the scale can be used You will find this information in Chapter 2 Please read these instructions carefully and completely...

Страница 5: ...Engage the lift mechanism by lowering the hand control It may be necessary to hit the release button before doing so 7 Lift the pallet by cranking the handle in an UP DOWN direction a minimum of 3 com...

Страница 6: ...ING 0 Better known as the Center of Zero annunciator this light is active whenever the displayed weight is within 0 25 divisions of true zero N Indicates that the scale is displaying net weight G Indi...

Страница 7: ...cale is not at or below Gross zero 2 The scale is not in motion 3 The scale is not in overload see Appendix C for error codes ON Print This key is used to turn the scale ON The Print function is not a...

Страница 8: ...he display shows zero weight and turns the NET annunciator on 5 Place the material to be weighed on the container or pallet raise it off the floor and allow the weight indication to stabilize 6 Read t...

Страница 9: ...ontainer or pallet you must first tare the container or pallet off by pressing the TARE key 3 If the screen does not show 10 0 press the TARE key once The scale is prompting you to place ten identical...

Страница 10: ...the switch in the down position as shown at right 4 Power on the indicator by pressing the ON key The indicator shows F 1 to indicate that you are in Setup Menu mode CANNED A D CONVERTER SETUP SWITCH...

Страница 11: ...ound in Table 5 1 Factory set defaults are shown in bold with a checkmark 5 2 4 EXITING THE USER MENU 1 Power off the indicator by pressing and holding the down ZERO OFF key for 3 seconds 2 Move the S...

Страница 12: ...vel begins the sequence 0 199999 123456 A6 No of line feeds Actuates the function that allows entry of the desired number of line feeds to be printed in the Print Ticket Mode Valid only when A3 is set...

Страница 13: ...e indicator must be in standstill to zero the scale 100 1 9 F5 Motion Band Sets the level at which motion is detected by comparing the present display update with the previous one If motion is not det...

Страница 14: ...Places indicator into the zero calibration routine Scrolling down with the ZERO key one level begins the procedure Press ZERO key to begin sequence F17 Span Calibration Places indicator into the span...

Страница 15: ...15...

Страница 16: ...e the zero point value The display will show EndC0 momentarily then revert back up to F16 At this time proceed to the F17 span calibration to complete indicator calibration 6 3 SPAN CALIBRATION F17 1...

Страница 17: ...Change the calibration test weight or check the input data Err1 The calibration test weight or the adjusted keyed in weight is smaller than 1 of the full capacity of the scale Change the calibration t...

Страница 18: ...Scale Capacity OPERATING TEMPERATURE RANGE 14 F to 104 F 10 C to 40 C POWER SOURCE 2 internal 6V lead acid batteries AC Adapter 12VDC 800 mA Also acts as charger CHARGER See Power Source above Charge...

Страница 19: ...eck all load connections Use F16 mode to view internal counts See Appendix C for more information Err 3 All Modes Non volatile memory read error One or more setup parameters have been lost Err 4 All M...

Страница 20: ...scale may be used for an additional 15 to 30 minutes without damage to the internal battery If the scale is used for more than 30 minutes after a low battery indication permanent damage may occur to t...

Отзывы: