CHAPTER 6: SCALE CALIBRATION

6.1

SCALE CALIBRATION OVERVIEW

The  PALLET  RUNNER  3000  scale  may  be  calibrated  with  any  precision  test  weight

between 30 pounds and 3000 pounds. The recommended test weight is 2000 pounds, but

in general the larger the test weight, the more accurate the scale will be.

Before you can calibrate the scale, you must enter the Setup Menu. See Section 5.2.1 for

instructions on how to do this.

6.2

ZERO CALIBRATION (F16)

1.  While in the Setup mode, scroll to "

F 16

", then scroll down once using the ZERO/OFF

key to enter zero calibration menu. The display will momentarily show "

C 0

" followed by

a  value.  This  value  is  the  internal  A/D  count  and  can  prove  useful  when  trying  to

troubleshoot setup problems.

2.  After making sure that there are no test weights on the platform, press the ZERO/OFF

key again to zero out the displayed value.

3.  Press the NET/GROSS key to save the zero point value. The display will show "

EndC0

"

momentarily,  then  revert  back  up  to  F16.  At  this  time,  proceed  to  the  F17  span

calibration to complete indicator calibration.

6.3

SPAN CALIBRATION (F17)

1. While in the Setup mode, scroll to "

F 17

", then scroll down once using the ZERO/OFF

key to enter span calibration menu.

2.  The  display  will  momentarily  show  "

C  1

"  for  the  span  calibration,  followed  by  a  value

with one flashing digit. Place the test weight on the weighing mechanism.

3.  Use the four directional keys (shown in Figure 6-1 below) to adjust the displayed value

to the actual test weight value. Increase the flashing digit by pressing the UNITS key.

Decrease  the  flashing  digit  by  pressing  the  ZERO/OFF  key.  Pressing  the  ON/PRINT

key or the TARE key will change the position of the flashing digit.

SETUP MODE KEY FUNCTIONS

 

ZERO 

GROSS 

NET 

TARE 

PRINT 

SET

 

UNITS 

OFF 

ON 

Figure 6-1: Setup Menu Key Assignments

4. After setting the exact value, press the NET/GROSS key to save the value.
5. If the calibration was successful, the display will show "

EndC1

" momentarily, then revert

back up to F17.

Содержание PALLET RUNNER 3000

Страница 1: ...ruary 13 2004 Contents subject to change without notice Salter Brecknell Weighing Products 1000 Armstrong Drive Fairmont MN 56031 Tel 800 637 0529 Tel 507 238 8702 Fax 507 238 8271 E mail sales salter...

Страница 2: ...n Item 7 3 3 2 Taring an Item 7 8 Chapter 4 Advanced Features and Operation 9 4 1 Piece Counting 9 Chapter 5 Scale Configuration 10 5 1 Scale Configuration Overview 10 11 5 2 User A Menu 12 5 2 1 Ente...

Страница 3: ...internal battery 20 D 4 How long to charge the internal battery 20 LIST OF FIGURES 1 1 PALLET RUNNER 3000 Front Panel 4 3 1 PALLET RUNNER 3000 Display Detail 6 3 2 Function Keys Layout 7 5 1 PALLET RU...

Страница 4: ...ed after unpacking the scale A small amount of preparation is necessary before the scale can be used You will find this information in Chapter 2 Please read these instructions carefully and completely...

Страница 5: ...Engage the lift mechanism by lowering the hand control It may be necessary to hit the release button before doing so 7 Lift the pallet by cranking the handle in an UP DOWN direction a minimum of 3 com...

Страница 6: ...ING 0 Better known as the Center of Zero annunciator this light is active whenever the displayed weight is within 0 25 divisions of true zero N Indicates that the scale is displaying net weight G Indi...

Страница 7: ...cale is not at or below Gross zero 2 The scale is not in motion 3 The scale is not in overload see Appendix C for error codes ON Print This key is used to turn the scale ON The Print function is not a...

Страница 8: ...he display shows zero weight and turns the NET annunciator on 5 Place the material to be weighed on the container or pallet raise it off the floor and allow the weight indication to stabilize 6 Read t...

Страница 9: ...ontainer or pallet you must first tare the container or pallet off by pressing the TARE key 3 If the screen does not show 10 0 press the TARE key once The scale is prompting you to place ten identical...

Страница 10: ...the switch in the down position as shown at right 4 Power on the indicator by pressing the ON key The indicator shows F 1 to indicate that you are in Setup Menu mode CANNED A D CONVERTER SETUP SWITCH...

Страница 11: ...ound in Table 5 1 Factory set defaults are shown in bold with a checkmark 5 2 4 EXITING THE USER MENU 1 Power off the indicator by pressing and holding the down ZERO OFF key for 3 seconds 2 Move the S...

Страница 12: ...vel begins the sequence 0 199999 123456 A6 No of line feeds Actuates the function that allows entry of the desired number of line feeds to be printed in the Print Ticket Mode Valid only when A3 is set...

Страница 13: ...e indicator must be in standstill to zero the scale 100 1 9 F5 Motion Band Sets the level at which motion is detected by comparing the present display update with the previous one If motion is not det...

Страница 14: ...Places indicator into the zero calibration routine Scrolling down with the ZERO key one level begins the procedure Press ZERO key to begin sequence F17 Span Calibration Places indicator into the span...

Страница 15: ...15...

Страница 16: ...e the zero point value The display will show EndC0 momentarily then revert back up to F16 At this time proceed to the F17 span calibration to complete indicator calibration 6 3 SPAN CALIBRATION F17 1...

Страница 17: ...Change the calibration test weight or check the input data Err1 The calibration test weight or the adjusted keyed in weight is smaller than 1 of the full capacity of the scale Change the calibration t...

Страница 18: ...Scale Capacity OPERATING TEMPERATURE RANGE 14 F to 104 F 10 C to 40 C POWER SOURCE 2 internal 6V lead acid batteries AC Adapter 12VDC 800 mA Also acts as charger CHARGER See Power Source above Charge...

Страница 19: ...eck all load connections Use F16 mode to view internal counts See Appendix C for more information Err 3 All Modes Non volatile memory read error One or more setup parameters have been lost Err 4 All M...

Страница 20: ...scale may be used for an additional 15 to 30 minutes without damage to the internal battery If the scale is used for more than 30 minutes after a low battery indication permanent damage may occur to t...

Отзывы: