11
Current setting,
Zero0.5
through
Zero5.0
, is displayed.
4.
Press the
F
key repeatedly to scroll through the choices. Choices
are in 1/2 division increments. When your choice is displayed, press
the
TARE
key…
5.
The function is set and the scale returns to normal weighing mode.
bUZZEr
Use this item to set the Accept buzzer or beeper on or off. If you choose
on, the unit will beep when a steady weight is in the Accept range. If you
choose oFF, the beeper will not sound.
Set using the directions below.
1.
Press and hold the
F
key for three seconds to access the
Configuration menu …
rAnGE
is displayed.
2.
Press the
F
key five times…
bUZZEr
is displayed.
3.
Press the
TARE
key…
Current setting,
on
or
oFF
, is displayed.
4.
Press the
F
key to toggle between the choices. When your choice is
displayed, press the
TARE
key…
5.
The function is set and the scale returns to normal weighing mode.
7.0
Calibration
You can calibrate this scale using single point or multpoint calibration.
Single point calibration uses just the empty scale and a test weights equal
to full capacity to calibrate the scale. During multipoint calibration, you can
choose between a powerup zero range of 20% or 100% and use zero
weight, 1/3 capacity, 2/3 capacity and full capacity test weights.
Single Point Calibraiton:
Follow these steps for single point calibration of the scale. You will need
test weights equal to full capacity.
1.
Place the calibration switch, located on the bottom of the scale into
the ON position. The ON position is when the switch is pulled
towards the front of the scale. You may need to remove the small
cover to access the calibration switch. Use a small screwdriver to
remove the two screws holding the cover.
Содержание C3235
Страница 15: ...15...