978-922-1832
●
www.sagescience.com
●
BluePippin Operations Manual 460013 Rev F
25-7
25.6
0.75% Agarose, 1-6 kb, Marker S1, Low Voltage
(external marker)
Use Product Number:
BLF7510
Use Cassette Definition:
0.75% Agarose Dye Free /
0.75% DF Low Voltage 1 - 6kb Marker S1
DNA Marker S1 - Typical times to detector
Estimated Collection Times for Minimum Size Range (Tight) Targets
* Run times are affected by temperature and broad range collections.
Times should be used as an approximate guideline.
Performance Specifications
Minimum Size Distribution as CV
10%
Sample Recovery
50
– 80%
Reproducibility
<10%
Specifications subject to change without notice
* see page 25-1 for specification definitions
Target
(bp)
Time to Collect*
(hr:min)
1,000
3:04
2,000
3:45
3,000
4:22
4,000
5:10
5,000
5:46
6,000
6:25