If no faults are found,
LED’s will turn off if no products are present in warmer slots. If products are present, the
LED’s for the filled slots will indicate “NOT READY.”
If a fault condition exists,
LED’s will not light up if product is present in warmer slots. Unused slots will indicate a
“NOT READY” status. Discontinue use of the warmer unit and contact Sage Products.
2.
To stock warmer,
open door(s) and fill slots with any combination of Comfort Personal Cleansing products—one
package per slot. Place packages on shelves so they do not interfere with closing of door(s). Packages should be
placed far enough back that they contact rear of shelves, but not beyond. Close warmer door(s).
DO NOT
place
previously opened packages into warmer.
3.
Packages require approximately 2 hours to warm.
Products are ready for use when the “READY” indicator is
illuminated. To remove product for use, open door(s). Remove package(s). Close warmer door(s), making certain
door(s) close completely.
4.
Replacement parts and service:
Before performing any repairs or service on this warmer, contact Sage Products.
5.
Fuses:
This warmer is equipped with two externally accessible fuses. Turn off and unplug the warmer before
attempting to remove fuses. If the warmer stops running, check the fuses on the back side of the cabinet’s exterior.
If a fuse is blown, replace it with another fuse of identical rating. If the new replacement fuse blows, contact
Sage Products for further instructions.
To replace fuse:
Turn off and unplug unit. Find the round knob on the rear panel of the warmer cabinet
labeled “FUSE” and unscrew the fuse cover counterclockwise as indicated by the arrow on the fuse cover.
Remove the fuse and replace it with an 8-Amp, 250-Volt (8A, 250V) fuse and replace the fuse holder
cover. Plug the warmer into the 115V power outlet and turn power switch to the “on” position. If warmer
will not run, or does not run properly after replacing the fuse, call Sage Products for further direction.
6.
Memory Backup:
This warmer is equipped with a
5-minute memory backup
. It provides backup power to the
status indicator system for a short period of time in the event that the power switch is shut off, or power cord is
removed from the wall outlet. If power is interrupted, the status indicators will remember the status of all the
packages in the warmer at that time and reinstate those settings when power is restored within 5 minutes.
7.
Comfort Personal Cleansing products:
Refer to individual product package for instructions, ingredients,
warnings and other information.
Front View
#7944 (14 count)
#7945 (28 count)
Side View
Front View
Side View