background image

This warmer is to be used SOLELY for Sage Comfort Personal Cleansing

®

products. The warmer has been 

designed and safety tested to be used exclusively with Comfort Personal Cleansing products. ANY other use of the
warmer by the facility, including but not limited to use with other bathing products, can result in overheating, fire or
other hazardous conditions and is expressly forbidden. 

WARNING:

To reduce the risk of fire, burns, electrical shock, or injury:

1.  Read all instructions before using this warmer. 

2.  Grounding reliability can only be achieved when the equipment is connected to an equivalent receptacle marked 

“hospital only” or “hospital grade.”

3.

Install this warmer only in accordance with the installation instructions found in this manual.

4. 

DO NOT

cover or block any openings on this appliance.

5. 

DO NOT

tamper with or make any adjustments to the control panel, housings, safety switches, or any other part 

of the Comfort Personal Cleansing Warmer.

6. 

DO NOT

remove any panel or cover.

7. 

DO NOT

store or use this appliance outdoors. 

DO NOT

use this product near water—for example, near a kitchen 

sink, in a wet basement, or near a swimming pool.

8. 

DO NOT

use this warmer for any material other than Comfort Personal Cleansing products in accordance 

with instructions on the label.

9. 

DO NOT

let the cord hang over edge of table or counter.

10.  This warmer is to be used by adults familiar with its operation and these safety instructions.

11. 

DO NOT

operate this appliance if it has a damaged cord or plug, if it is not working properly, or if it has been 

damaged or dropped.

12. 

DO NOT

immerse cord or plug in water.

13.  Keep cord away from heated surfaces.

14.  The warmer is to be serviced only by qualified service personnel. If there are no external or user-serviceable 

fuses, contact Sage Products for repair or adjustment. If there 

are

external or user-serviceable fuses, check them and 

replace if necessary. If new fuses do not correct the condition, or if fuses continue to blow, contact Sage Products 
for repair or adjustment.

15. Use this appliance only for its intended purpose as described in this manual. 

DO NOT

use corrosive chemicals 

or vapors in this appliance. This type of warmer is specifically designed to warm Comfort Personal 
Cleansing products. It is 

NOT

designed for industrial or laboratory use. The use of corrosive chemicals in heating 

or cleaning will damage the appliance and may result in product failure.

IMPORTANT SAFETY INSTRUCTIONS

Содержание Comfort 7944

Страница 1: ...Shock Type B Applied Parts Classification Class I Y9891B Sage Products Inc 2003 Degree of Protection Against Ingress of Water Ordinary Mode of Operation Suitable for Continuous Operation Storage Tempe...

Страница 2: ...4 14 count Model 7945 28 count Please read and understand these instructions completely prior to operating the Comfort Personal Cleansing Warmer For use with Comfort Personal Cleansing brand bathing i...

Страница 3: ...near a swimming pool 8 DO NOT use this warmer for any material other than Comfort Personal Cleansing products in accordance with instructions on the label 9 DO NOT let the cord hang over edge of tabl...

Страница 4: ...the warmer long enough to reach the target temperature and is ready for use 3 TAKE FIRST The blinking green arrow and message that reads TAKE FIRST indicates the package in this slot has been in the w...

Страница 5: ...e electrical outlet d Be certain electrical outlet is not blocked by the warmer Power cord must be able to be unplugged without moving the warmer Grounding Instructions This appliance must be grounded...

Страница 6: ...sively hot do not use Remember gloves diminish your sensitivity to heat Closely monitor heat level with infants unresponsive or skin sensitive individuals Caps Baths are for individual use only This w...

Страница 7: ...emove fuses If the warmer stops running check the fuses on the back side of the cabinet s exterior If a fuse is blown replace it with another fuse of identical rating If the new replacement fuse blows...

Страница 8: ...umns and center column must be removed to allow cleaning of the shelves and cabinet interior Remove the Phillips head screw located at the base of each column Gently lower the column to disengage from...