Upon coming across any problem in operation, follow the instructions below to check and
repair the problem, and/or get in touch with your representative.
UNFORESEEN EVENTS – SOLUTION OF PROBLEMS
Imprevistos
Provável Causa
Soluções
Problem
Probable cause
Solution
- The Curing Light does not
work.
- Battery in the handpiece
without charge.
- Overheating protection
activated “error code: oU.
- LED damaged “error
code: Er.
- Recharge the handpiece on
the base for 3 hours.
- Wait a few minutes.
- Get in touch with the Alliage
technical assistance.
- The equipment is not
polymerizing the resins.
- Resin not appropriate for the
wavelength range of the LED
curing lights.
- Light conductor fastened
incorrectly.
- Residue of resin in the light
conductor.
- L i g h t c o n d u c t o r w i t h
protection cover.
- Acquire the appropriate resin
for the wavelength of the
curing light, i.e., which contain
p h o t o i n i t i a t o r s w i t h
canforoquinone.
- Fasten the light conductor
correctly.
- Clean the light conductor.
- Remove the protection cover
from the light conductor.
- Inadequate luminous power. - Light conductor fastened
incorrectly.
- Problems with the light
conductor.
- Reduced capacity of the
battery.
- Fasten the light conductor
correctly.
- Replace the light conductor.
- Get in touch with the Alliage
technical assistance.