24
CORRECTIVE AND PREVENTIVE MAINTENANCE AND PRESERVATION
Attention
Cleaning, disinfection, and sterilization have limited impact on the reusable parts of the
equipment. Therefore, the number of repeated procedures should not exceed 20 times.
Cleaning ,Disinfection and Sterilization
Preparation for use
Immediately after use, the reprocessing equipment is immersed in <40 °C tap water (drinking water quality,
‘water' mentioned in this chapter, which is required to meet this standard) to remove dirt. Do not use a
fixed detergent or warm water (>40 ° C) as this will cause the residue to be fixed and affect the effect of
repeated treatment.
Transport: Transport to the post-processing area for safe storage to avoid any damage and environmental
pollution.
Preparation before cleaning
Remove the body light guide from the main unit and place it in a stainless steel box.
Decontamination preparation (pre-cleaning):
Rinse the body light guide with running tap water (<40 °C) until all visible residue is removed.
Manual cleaning
Rinse the body light guide with running tap water (<40 ° C) and gently wipe the visible dirt with a soft brush;
Place the body light guide in a multi-enzyme cleaner and soak for 10 minutes to decompose the
contaminants. Follow the instructions of the cleaning agent manufacturer;
The body light guide was placed under running tap water for at least 1 minute to remove the residue of
the cleaning agent.
Attention
We recommend the use of proven Lilcon® multi-enzyme cleaning solutions or multi-enzyme
cleaning solutions that comply with local regulations (eg CE, FDA approval).
Manual disinfection
Place the body light guide in a dish containing the cleaning and disinfecting solution for 10 minutes for
immersion disinfection.
The body light guide was rinsed with running tap water (<40 ° C) for at least 1 min to remove residuals
from the disinfectant.
Disinfectant: It is recommended to use Ronso O-Benzaldehyde Disinfectant (OPA), no need to match.
Attention
1) If other disinfectant is used, use a disinfectant that complies with local national
regulations (such as CE certification, FDA certification), and follow the instructions provided
by the manufacturer of the disinfectant.
2) In order to prevent the disinfectant from corroding the device, the time of immersion
disinfection should not be too long (<30min).