7. Summary Table of the NEG Pumps controlled by the NEG POWER
GP PUMPS
Model Pump
Notes
Activation
Voltage (nude)
[V]
Activation
Current (nude)
[A]
Activation
Power (nude)
[W]
Temperature
[°C ]
GP 50-2F
st 707
25
2.8
75
(450°C) NO TCK
GP 50-2F
st 101
49
4.6
225
(700°C) NO TCK
GP 100
st 707 - MK5
41
3.4
138
450
GP 100
st 101 - MK4
81
5.2
420
700
GP 200
st 707 - MK5
52
3.8
200
450
GP 200
st 707 - MK4
25
7.0
175
450
GP 500
st 707 - MK5
61
9.4
580
450
GP 500
st 707 - MK4
46
9.9
455
450
GP 500 - SP8
st 707
110
4.5
500
500
GP 500-2X
st 101 - MK5
70
10
700
700
CAPACITORR PUMPS
Model Pump
Notes
Activation
Voltage (nude)
[V]
Activation
Current (nude)
[A]
Activation
Power (nude)
[W]
Temperature
[°C ]
CT D400-2
st 172
16
5.0
90
(450°C) NO TCK
CT D1000
st 172
43
4.3
190
550
CT D2000
st 172
110
4.6
500
450
CT D3500
st 172
92
3.9
360
550
NEXTORR PUMPS
Model Pump
Notes
Activation
Voltage (nude)
[V]
Activation
Current (nude)
[A]
Activation
Power (nude)
[W]
Temperature
[°C ]
NXT 100-5
st 172
9
5.0
45
(450°C) NO TCK
NXT 200-5
st 172
12.5
4.8 60
(450°C)
NO
TCK
NXT 300-5
st 172
20
4.6
92
550
NXT 500-5
st 172
24
4.8
120
550
NXT 1000-10
st 172
58
4.5
260
550
NXT 2000-10
st 172
66
4.2
280
550