3.  INSTALLATION 

The  unit may  be  installed  on  a  desk  or mounted  in  a rack.  Place the  instrument  away  from  heat 
sources. A gap of at least 5 cm must be left between the sides of the unit and adjacent walls to 
ensure sufficient air flow to the cooling fan. If the NEG POWER is installed inside a cabinet, the latter 
must be provided with vents capable of ensuring adequate air exchange between the inside and the 
outside. 

 

3.1. 

Rack mounting 

Insert the instrument in the appropriate space (19” rack 3U according to DIN 41494 standard). 

Fix the NEG POWER MINI front panel on the rack with four screws. 

 

 ATTENTION:

 the instrument cannot be used in potentially flammable or explosive areas. 

 

  ATTENTION: 

to  avoid  damages  of  the  electronic  circuit  use  the  instrument  within  the 

environment specified conditions.  

 

  ATTENTION: 

do  not  use  accessories  which  are  not  supplied  or  recommended  by  the 

manufacturer. 

 

 

3.2. 

Electrical connections 

The connections for the pump, the mains line, the interlock and the external interface are on the rear 
panel of the instrument. 
Before connecting the unit to the mains supply, make sure that supply voltage is within the range 
specified in the section Electrical Specifications. 
 

  ATTENTION:

  protective earthing is required! 

Plug the equipment into a supply outlet which has an earth connection. 

 

 

3.3. 

Cleaning 

 ATTENTION:

 before starting any cleaning turn the unit off and remove all mains. 

 
The units should be cleaned with a damp cloth when necessary.  
The use of solvents or of an aggressive or scouring cleaning agent is not permitted.  
Avoid spraying or spilling any type of liquid directly onto the unit.  
Before putting the unit into operation mode, check that it is thoroughly dry. 
 

 

 

Содержание NEG Power mini

Страница 1: ...UDOC CONTROLLED COPY M HIST 0138 23 R1 October 06 th 2020 SAES Getters S p A Italy www saesgroup com USER MANUAL NEG POWER MINI NEG Pump Controller...

Страница 2: ...least 3 hours at the ambient temperature WARNING The use of the instrument for any other purpose than the one originally intended may damage the instrument injure the user and void all warranties ATT...

Страница 3: ...ORMAL SITUATION AND ALARM NOTIFICATION 13 5 1 INTERLOCK Alarm 13 5 2 PUMP OPEN Alarm 14 5 3 OVER CURRENT Alarm 14 5 4 PUMP OVER TEMPERATURE Alarm 14 5 5 VMON UNDER VOLTAGE OVER VOLTAGE 15 6 RECOMMENDE...

Страница 4: ...rs on a USB flash memory The controller case is a sub rack 3U 28TE 230mm deep output connector on the back takes 30 mm of additional space room for male connector should also be accounted for For the...

Страница 5: ...on the back the instrument input for mains power supply IEC 10A plug with fuse an ON OFF button to turn on off the instrument one OUTPUT pump connector a fan IN OUT interface connector with INTERLOCK...

Страница 6: ...play 2 MENU ENTER button to edit the output parameters and to enter the setup menu 3 to reduce the selected value scroll down in the menu 4 to increment the selected value scroll up in the menu 5 STAR...

Страница 7: ...OUT interface connector with INTERLOCK and RS232 connector 11 ON OFF instrument power switch 12 instrument mains plug 2 6 NEG POWER MINI Dimensions and Weight NEG POWER MINI dimensions 28TE 19 sub rac...

Страница 8: ...onic circuit use the instrument within the environment specified conditions ATTENTION do not use accessories which are not supplied or recommended by the manufacturer 3 2 Electrical connections The co...

Страница 9: ...witch into position I The display turns ON Note The NEG pump is not electrically powered when the supply is switched on The first screen shows the name of the instrument and the software version After...

Страница 10: ...ressing the button once starts supply and pressing it again stops the supply at any time START STOP is not enabled while editing parameters When you turn off and turn on NEG POWER MINI the last settin...

Страница 11: ...page shows the software version the hardware version and the MAC address With the down arrow you can scroll to the second page Here MODBUS TCP can be enabled and IP parameters can be changed A short...

Страница 12: ...wing parameters are available Enable Disable Enable Disable the Modbus over RS232 Baudrate is the serial port speed in bit s Parity is the parity bit configuration of RS232 Stop is the stop bit config...

Страница 13: ...er values The setting values are also reported in the file every time they are changed See the following example t s Vout V Iout A Tpump C Vset 24 0 Trise 000 Thold 000 0 23 9 5 2 26 10 24 0 5 2 26 20...

Страница 14: ...the alarm persists please contact SAES GETTERS S p A for assistance 5 3 OVER CURRENT Alarm This notification happens when the output current is permanently over 5A It indicates that the electrical se...

Страница 15: ...E OVER VOLTAGE Other possible alarm situations are VMON or UNDER VOLTAGE or OVER VOLTAGE These notifications indicate that there are some problems with the internal circuit and or with the internal re...

Страница 16: ...neral indicative rule is to apply the maximum voltage during activation of the pump in the nude configuration without jacket If the pump is with jacket voltage to be applied should be reduced by 10 20...

Страница 17: ...AME Vmax Vmax CT D100 9 V 3 5 V CT D200 12 V 4 V CT D400 2 16 V 7 V CT D50 8 V 3 V CT HV 200 35 11 3 V 3 3 V CT HV 200 63 12 6 V 3 7 V CT Z100 8 V 4 V CT Z200 11 V 5 V CT Z400 body 15 V 5 V CT Z400 nu...

Страница 18: ...ble to reduce Trise down to 0 in this case Vset will be reached in a few seconds Once Vset is reached Vset voltage is kept for Thold 6 3 Heating without automatic stop Some applications require specia...

Страница 19: ...be used for communication purposes via RS232 interface in MODBUS RTU protocol LAN connector can be used for communication purposes via ETHERNET interface in MODBUS TCP protocol Modbus Application Pro...

Страница 20: ...up to 2000 m Operating Temperature 0 C to 40 C Relative Humidity 80 Noise at 1 mt 40 dBA Output Output power 175 W overload Output Voltage 0 35 V Supply Current 5 A Overload 110 9 MAINTENANCE The supp...

Страница 21: ...for MK4 type pumps 3B0345 3B0356 3B0353 Cable supply output for MK5 type pumps 3B0337 3B0361 3B0364 and for CapaciTorr D2000 and D3500 pumps Cable supply output HT for MK5 type pumps 3B0393 3B0394 3B0...

Страница 22: ...2 19 EU directive Waste Electrical and Electronic Equipment The instrument contains materials which may endanger the environment and it s not allowed to dispose it with unsorted urban waste The equipm...

Страница 23: ...thirty 30 days of BUYER s discovery of the breach This warranty shall only apply to the BUYER and may not be assigned During the term of the warranty set forth above SAES will promptly repair the Pro...

Страница 24: ...2 Fax 39 02 93178320 E mail CRM_SALES saes group com Asia and Oceania SAES Getters S p A Japan Technical Service Branch Office 2nd Gotanda Fujikoshi Bldg 23 1 Higashi Gotanda 5 Chome Tokyo 141 Japan P...

Страница 25: ...Green 2 1 gnd 3 5 Yellow Green Blue Brown TO NEG PUMP CONTROLLER REAR VIEW OF CONNECTOR TO GETTER PUMP CapaciTorr D400 2 CapaciTorr Z400 CapaciTorr HV200 35 CapaciTorr HV200 63 For these pumps which a...

Страница 26: ...with a dedicated electrical connector use the output supply cable shown in figure NexTorr D300 5 NexTorr D500 5 NexTorr HV 300 NexTorr Z300 NexTorr D500 StarCell For these pumps which are equipped wit...

Страница 27: ...M HIST 0138 23 R1 Page 27 of 28 October 06 th 2020 BLANK PAGE...

Страница 28: ...ST 0138 23 R1 Page 28 of 28 October 06 th 2020 SAES GETTERS S p A Italy www saesgroup com This document is property of SAES Group S p A The information contained herein is subject to change without no...

Отзывы: