THE AUTOSCALER DIRECTIONS FOR USE Rev: A March 16, 2007 

 

WHAT TO DO IF UNIT STOPS WORKING

 

 

If the unit does not function, proceed with the following: 

 

1. 

If the power L.E.D. light above the switch is not lit, replace the

 FUSE

 on 

the rear panel

 with a 

"LITTLE FUSE“2 amp Fast Blow style.

 If the fuse 

opens again 

DO NOT

  attempt to replace the fuse as this indicates a 

serious condition that must be examined by the factory. Call for technical 

service 

(519) 579-9323

  or ask your local dealer to look up our current 

phone number and address in their records, or the 

Proofs Magazine

 

listing of dental manufacturers. 

2. 

Check the plug.

 Is it all the way in the receptacle? 

3. 

Try another insert

. (Make sure it is the correct frequency 

25K or 30K

 

depending on your model.) If you have a 30K model a 

30K sticker

 will be 

on the front panel in the lower right corner or marked on the back of the 

unit usually on the serial number plate. 

The most common mistake 

when purchasing a new unit is selecting the wrong frequency 

Cavitron insert. 

4. 

If water flow is 

insufficient,

  check water pressure at the operatory for 

minimum of 

20 PSI

. The 

Insert Water SLOT

  could be clogged with 

debris. Try another insert. See if 

WATER  QUICK CONNECTOR

  is 

locked fully

 in place. 

5. 

If the handpiece handle or insert operates 

HOT:  REDUCE POWER 

LEVEL

 and 

INCREASE WATER FLOW. 

6. 

NEVER 

touch the Cavitating/Vibrating end 

of the insert tip to any 

FLESH

 such as the 

cheek, tongue, or gums unless performing special periodontal 

procedures. This will cause 

a friction burn

  to the patient. Always keep 

the insert in motion. Do not rest the handle or insert on the lips during a 

procedure. The 

REASON

  for this is due to the 

25,000  or 30,000 back 

and forth strokes of the insert every second.

 This is the principal idea 

of ultrasonic cleaning activity. The insert tip should touch 

PRIMARILY 

the 

calculus on the tooth surface, and NOT the enamel of the tooth surface. 

7. 

If 

THE AUTOSCALER

 

stops functioning or is intermittent 

when the 

handpiece cord is moved, this means a wire in the handpiece cord is 

broken and must be immediately repaired or replaced. 

DO NOT

 continue 

to press the foot switch as there is a possibility the internal circuit 

safety 

fuse

 may blow and require replacement by a qualified technician.

  

 

 

Содержание 30K

Страница 1: ...LTRASONIC SCALING DEVICES AUTOSCALER U L T R A S O N I C D E N T A L S C A L E R DIRECTIONS FOR USE DFU PLEASEREADTHOROUGHLYBEFOREOPERATINGTHISDEVICE HEINTENDEDUSEOFTHEAUTOSCALER DEVICEISFORDENTA OPHY...

Страница 2: ...e not interchangeable The Autoscaler accepts all major brands of scaling tips THE AUTOSCALER name is owned by Sable Industries Inc of Canada THE AUTOSCALER is acceptable for use in scaling and cleanin...

Страница 3: ...e is subject to deterioration by aging chemicals in the water supply excessive flexing and pressure fluctuations As an added precaution the water supply should be turned OFF at the end of the day REME...

Страница 4: ...p contamination from getting on the handpiece assembly Dispose sheathe after each patient Ask your local dealer for HIGH SPEED LONG sheathes INTRODUCTION THE AUTOSCALER Ultrasonic Scaler is a precisio...

Страница 5: ...Switch is set to the ON position indicating the unit is ready for operation POWER CONTROL KNOB NOTE FOLLOW INSERT MANUFACTURERS COMPLETE INSTRUCTIONS Linear Power Control knob adjusts the output power...

Страница 6: ...ating at a higher than normal temperature and should be replaced immediately NOTE Insert tips manufactured by other manufacturers may not achieve optimal results and could run at a higher temperature...

Страница 7: ...ceipt of your unit carefully unpack it and see that the Toggle Switch is in the OFF O position 3 The Water Control Knob should be in the HIGH position Note It is normal for the Knob to turn past the H...

Страница 8: ...g Vibrating end of the insert tip to any FLESH such as the cheek tongue or gums unless performing special periodontal procedures This will cause a friction burn to the patient Always keep the insert i...

Страница 9: ...ion of the POWER KNOB is for controlling the Output Power to the insert tip in a smooth linear action The operator can select the EXACT power needed for cleaning Set power knob LOW to MEDIUM for coole...

Страница 10: ...water flow is insufficient check water pressure at the operatory for minimum of 20 PSI The Insert Water SLOT could be clogged with debris Try another insert See if WATER QUICK CONNECTOR is locked ful...

Страница 11: ...the Blue area at assure best performance Call your DEALER or customer service at Sable Industries Inc for further instructions at area 519 579 9323 and a technician will be available to help you with...

Страница 12: ...ing and handling back to the customer 4 UNAUTHORIZED REPAIR ABUSE ETC The unit must not have been previously altered repaired or serviced by anyone other than a service facility authorized by SII To r...

Страница 13: ...PT 5 GLOVES 3 H HANDPIECE ASSEMBLY 4 HANDPIECE INSTALLATION 5 HANDPIECE OR INSERT RUNS HOT 9 HANDPIECE PANEL RECEPTACLE 4 HANDPIECE SHEATHES 3HIGH POWER 8 HUMMING OR HIGH PITCH SOUND 10 I INSERT REPLA...

Страница 14: ...Page Number 1 THE AUTOSCALER DIRECTIONS FOR USE Rev A March 16 2007 SABLE ULTRASONIC TIPS 1900601 DF3 30K 1900602 DF3 25K 1900603 DFK10 30K 1900604 DFK10 25K...

Страница 15: ...TAIRES FORM S L UTILISATION DES DISPOSITIFS ULTRASONIQUES DE D TARTRAGE DOIVENT UTILISER CET APPAREIL VEUILLEZ LIRE AU COMPLET AVANT D UTILISER CET APPAREIL L AUTOSCALER EST CON U POUR LES TRAITEMENTS...

Страница 16: ...rentes ne sont pas interchangeables L Autoscaler est compatible avec tous les embouts de d tartrage des principales marques Le nom L AUTOSCALER est la propri t de Sable Industries Inc of Canada L AUT...

Страница 17: ...us l effet de l ge des produits chimiques contenus dans l eau d une flexion excessive et des changements de pression Par mesure de pr caution suppl mentaire il est pr f rable de FERMER l arriv e d eau...

Страница 18: ...TABLES POUR PI CE MAIN afin d viter la contamination de l accessoire Changez de gaine apr s chaque patient Communiquez avec votre distributeur pour obtenir des GAINES LONGUES POUR HAUTE VITESSE INTROD...

Страница 19: ...osition ON I L appareil est alors pr t fonctionner Le R GLAGE DE LA PUISSANCE REMARQUE RESPECTEZ LA LETTRE LES INSTRUCTIONS DES FABRICANTS D INSERTS Le bouton r gle la puissance de sortie de mani re l...

Страница 20: ...R EFFICACIT ou fonctionneront une temp rature anormalement lev e Dans ce cas ils doivent tre remplac s imm diatement REMARQUE Il est possible que les pointes d insert provenant d autres fabricants don...

Страница 21: ...ctu toutes les tapes suivantes 2 Lors de la r ception de votre appareil d ballez le avec pr caution et v rifiez que l interrupteur d alimentation est bien en position OFF O 3 Le bouton de r glage du d...

Страница 22: ...pour augmenter le d bit Un peu d AIR s vacuera Cette tape est IMPORTANTE L air emprisonn peut entra ner la formation de ZONES DE SURCHAUFFE reconnaissables un son per ant pouvant compromettre l effic...

Страница 23: ...vers la droite s il y a absence de brouillard ou pr sence d un tr s l ger brouillard Si vous utilisez des inserts SLIMLINE MC FSI MC ou des inserts parodontaux modifi s positionnez le bouton R GLAGE D...

Страница 24: ...e c t de l interrupteur ne s allume pas remplacez le FUSIBLE situ dans le panneau arri re par un PETIT FUSIBLE fusion rapide de 2 A Si le fusible grille nouveau N ESSAYEZ PAS de le remplacer tout lais...

Страница 25: ...n doit imm diatement tre r par ou remplac NE CONTINUEZ PAS appuyer sur la p dale En effet il est possible que le fusible de s ret du circuit interne soit grill et doive tre remplac par un technicien q...

Страница 26: ...ACHAT D ORIGINE pour se pr valoir de la garantie La garantie N INCLUT PAS les frais de manutention et d exp dition au client ou au fabricant Consultez les termes de la garantie pour prendre connaissa...

Страница 27: ...tion abusive ou contraire aux instructions La pi ce main ne sera pas garantie si elle a fait l objet d une utilisation abusive impliquant une traction ou un tirement excessifs ou encore l accumulation...

Страница 28: ...E FUITE LA TERRE 5 GANTS 3 H PI CE MAIN 4 INSTALLATION DE LA PI CE MAIN 5 PI CE MAIN OU INSERT QUI SURCHAUFFENT 9 SUPPORT DE LA PI CE MAIN 4 GAINES POUR PI CE MAIN 3 PUISSANCE LEV E HIGH 8 BOURDONNEME...

Отзывы: