THE AUTOSCALER DIRECTIONS FOR USE Rev: A March 16, 2007
HANDPIECE ASSEMBLY
The handpiece assembly is used to accept all brands of insert tips. It
is connected to
THE AUTOSCALER
front panel by a Quick Connector. The
open end
of the
handpiece will accept all brands of inserts, Inserts are
frequency specific.
!
30K tips measure (Shorter style approx. 6.25 inches long).
!
25K tips measure (Longer style approx. 7.25 inches long).
(25K) inserts will
NOT work in your
30K unit.
(30K) inserts will
NOT work in your
25K unit.
As illustrated both
frequency inserts look identical except for length.
(NOTE):
Inserts are extremely
FRAGILE
. If one is dropped, bent, or
scratched it
WILL NOT
cavitate
(I.E. will non function)
or function by
operating at a higher than normal temperature, and should be replaced
immediately
.
(NOTE):
Insert tips manufactured by other manufacturers may
not achieve optimal results, and could run at a
higher temperature or not
tune efficiently
in
THE AUTOSCALER
.
Insert tips should be routinely
replaced according to insert manufacturers recommendations.
ELECTRICAL REQUIREMENTS
A
grounded
wall receptacle is required supplying
a voltage range
of
110-125 volts 50/60 Hz A.C.
, or
(optional voltage range of) 220-240
volts 50/60 Hz A.C.
If a 3 to 2 prong adapter is used, the ground wire
MUST
be connected to the base plate screw.
Have an electrician test your outlet
for correct wiring polarity, and a safe ground. Do not assume the ground is
complete.
If you plug THE AUTOSCALER in a GFI (Ground Fault
Interrupt) outlet it may trip the socket due to the high sensitivity of the
socket. If this occurs plug THE AUTOSCALER into a non-GFI socket.