8-3 Preventive Maintenance (PM)
To avoid damage to the equipment, disinfection should be performed according to the
Hospital Maintenance Schedule.
Daily check of the following items is recommended:
1. Accessories intactness (no mechanical damage)
2. Accessories function
Weekly check of the following items is recommended:
1. The system cleanness
2. The system intactness (case, screen, keys, indicators, recorder door and release button)
3. The recorder function
Monthly check of the following items is recommended:
1. Calibration label (Send the system to the manufacturer for calibration on the predetermined
date).
2. The system intactness (no mechanical damage)
3. The system cleanness
4. Keys and indicators function
5. Accessories intactness (no mechanical damage)
6. The recorder function
It is recommended that the system is calibrated by manufacturer every
year, but it has to be calibrated once every 2 years. In addition, the system
lifetime is 10 years.
The medical center can request the system calibration whenever the
system accuracy is in doubt.
.