background image

700-3000  6.21.2016 

8

the service light is illuminated the lift must not be used until required maintenance is 
performed.

   

 
Emergency Stop Button –  
 

In the event of an emergency, or if you need to stop lift movement immediately, pressing the 
Emergency Stop Button (red button on the control) will stop all lift movement. At the same time an 
audible alert will sound for 10 sec. then pause for 5 sec. and then the sequence will repeat until 
the button is reset by turning it in the direction of the arrows on the button (clockwise). The 
Emergency Alert LED will also flash Red when the button is pressed and will remain flashing until 
the button is reset. 
 
The Emergency Alert (audible and LED) can be activated by pressing any two buttons on the 
hand control at the same time (provided with the lift). This will stop all lift movement and activate 
the audible alert and the Red LED will flash.  Once both buttons are released, the lift will return to 
normal operation and the emergency audible alert will silence.    

 
Activation Key – 

 

The Optional Activation Key provides a higher level of security by 
allowing control operation only with the use of the optional activation 
key. To enable operation slide the Access Key into the slot next to the 
touchpad control arrows. To disable operation remove the key. When 
the key is inserted or removed the Battery Level LED will illuminate for 
10 seconds. 
 

USB Port – 

 
The LiftOperator control contains a USB port that is used to download the performance log of the 
lift. Using any standard FAT file formatted USB memory stick (widely available through retail 
outlets) insert the stick into the slot on the bottom of the control on the far right hand side. Press 
any of the touchpad control arrows; all three of the Battery Level LED’s will flash. During 
downloading the Green greater than 50% Battery Level LED will flash. If the unit has a problem 
downloading the performance log the Red less than 25% LED will flash.  Remove the USB from 
the port and re-insert it.  If the Red 25% LED flashes again, confirm that the USB you are using is 
FAT file formatted. The information from the USB can be downloaded to a PC for review and/or 
copied to a spread sheet. The performance log will retain approximately 7,600 events. Once the 
file is full it will remove the oldest event and replace with the newest.    
 

After the summary data, the events are shown with the oldest first.  The first column is the time in 
seconds from that last event.  The second column is the text describing the event.  The third 
column is duration of the activity noted in seconds.

 

 

 

 

 

 

An example of a short log file: 

 

Activity   

 

 

Description 

dev_id_upper,0  

Programing data

 

dev_id_lower,1  

Programing data

 

 

fw_rev,90  

 

Firmware rev.

   

batt_volt_float,24929  

Non-loaded battery voltage in millivolts 

 

   Activation Key (Option)

Содержание aXs Series

Страница 1: ...00 3000 6 21 2016 1 aXs POOL LIFT S R SMITH LLC CORPORATE HEADQUARTERS P O Box 400 1017 S W Berg Parkway Canby Oregon 97013 USA Phone 503 266 2231 Fax 503 266 4334 OWNER S MANUAL MAINTENANCE PROCEDURE...

Страница 2: ...efore installing the pool lift the installer must review and confirm the information provided on the Deck Profile Sheet If the description of the application does not match the installation site a new...

Страница 3: ...IFT 12 TRANSFERRING 13 IN CASE OF BATTERY FAILURE Error Bookmark not defined STANDARD ACCESSORIES OPTIONAL ACCESSORIES 15 MAINTENANCE and CLEANSING 16 TROUBLE SHOOTING 17 LONG TERM STORAGE 18 WARRANTY...

Страница 4: ...ber of persons required to be in the water ready to receive the User The correct seat belt must be attached to the seat and fully fastened and used during each transfer Our goal is to provide our cust...

Страница 5: ...eck a minimum of 16 from the edge of the pool 1009 2 3 Clear Deck Space On the side of the seat opposite the water a clear deck space shall be provided parallel with the seat The space shall be 36 wid...

Страница 6: ...This design ensures consistent safe operation and minimizes service problems The aXs has a mast height of 36 92 cm above the deck provides 360 rotation and is intended to be used with simple pool gut...

Страница 7: ...is for the hand control Connector 1 is for the motor cable red stripe Connector 2 is for the actuator cable green stripe Touchpad Control The touchpad control arrow keys can be used in the event that...

Страница 8: ...en the key is inserted or removed the Battery Level LED will illuminate for 10 seconds USB Port The LiftOperator control contains a USB port that is used to download the performance log of the lift Us...

Страница 9: ...ck The battery pack is located on top of the Control Box and is removable To remove pull the battery pack away from the mounting plate so that the latch on the battery pack clears the side tabs on the...

Страница 10: ...e shortened Battery Disposal The batteries located inside the battery pack are recyclable and shall be disposed of in accordance with applicable local state provincial or federal national regulations...

Страница 11: ...arefully remove plastic covering from pallet 2 Carefully remove any wood bracing from pallet 3 Remove seat assembly and set aside for attachment 4 Remove lift from pallet and place into deck anchor 5...

Страница 12: ...d a risk assessment carried out If a User is mentally challenged or has severe physical disabilities these issues must be taken into account to determine the number of persons required to complete the...

Страница 13: ...f necessary If user has a wheelchair keep the wheelchair close by for easy retrieval Fasten Seat Belt thread loose end of belt strap through buckle pull tight to close press latch down on belt materia...

Страница 14: ...h with any questions concerning deck anchor placement or if the aXs will work in your installation Installing a Locking Anchor in a Concrete Deck The proper installation of the anchor depends on the c...

Страница 15: ...he lift to be turned manually Follow the proceeding process in reverse to reengage the gear assembly If the aXs will be used by a disabled person living on their own a communication device should be i...

Страница 16: ...schedule shall be performed to insure proper operation with the Daily items performed before each use Maintenance Performed Daily Weekly Monthly Check battery level before each use Charge battery dai...

Страница 17: ...ement information Does lift raise or lower No 1 Check battery charge level 2 Check battery connection 3 Use another fully charged battery If lift continues to not function replace the Control Box Lift...

Страница 18: ...lso include one photo or more of the damaged area 2 Write down or take a photo of the serial sticker from the product S R Smith provides a serial for every pool lift we produce The sticker with the se...

Страница 19: ...tage 24v DC Maximum Amp Draw 9 Maximum Speed 0 59 inch sec 1 49cm sec 3 Motor Rotation 24 VDC 13 RPM Gearing Ratio 9 1 4 Battery Power 24 VDC IP65 Gel Lead Acid Temperature Range 41 F 5 C to 104 F 40...

Страница 20: ...800 5065 Small Gear 18 800 5066 Toggle Clamp 19 AX4000 aXs Housing Back 20 AX4050 aXs Housing Front 21 AX9004 aXs Battery Console Cover Green not shown 22 AX9003 aXs Seat Belt Assembly Gray 23 AX8360...

Страница 21: ...700 3000 6 21 2016 21 Appendix A...

Страница 22: ...700 3000 6 21 2016 22 Appendix B...

Отзывы: