con::cube Network Integration and Security

For normal operation, remote control, data transfer and other features it might be necessary to setup a connectivity of the 

con::cube. Within this section the different possibilities are explained and security background is mentioned also.

The con::cube is based on Microsoft Windows XP embedded (XPe) operating system until moni::tool versions V2.5. From 

Version 3.0 on it is based on a Linux operating system (Debian).

While the XPe platform is well-proven and stable it already 

has reached the end of its mainstream support. Microsoft 

will still supply security updates within a certain period but 

won‘t fix non-critical issues. Linux will support further security 

updates.

Nonetheless,  the  con::cube  is  a  complex  device  offering 

services over the network. Software security therefore is an 

issue which needs to be considered to safely operate the 

con::cube.

On the con::cube the TCP / IP interface can be used via LAN, 

WLAN or 3G Modem (optional). All these features can be 

configured within the 

Service / Terminal / Network

 menu (see 

manual moni::tool).

The table below contains all types of possible connectivity setups. Regarding the individual connection types please refer to 

the subsections afterwards.

con::cube

Type of data transfer

Connection type

Customer infrastructure

Security

acts as client (i.e. con:: 

cube establishs the 

connection, performs 

a task and shuts down 

the connection)

  FTP data transfer

  SSH data transfer

  NTP-client (for system 

time synchronization)

all connection types

FTP server

SSH server

acts as server

(i.e. con::cube waits for 

a client to connect and 

responds to its request)

  Modbus TCP

 TML

  VNC remote control

  FTP remote control

LAN or WLAN

trusted network

3G private IP

internet

3G private IP

VPN server

3G public IP

open WLAN without 

WEP

Because of the number of services offered, the con::cube MUST NOT be operated in environments where those 

services can be reached public. i.e. the con::cube MUST NOT be reachable from the internet. The most common 

configurations that are NOT safe are a 3G connection with a public IP address and a WLAN connection to a public, 

non-encrypted network.

It is always safe to operate the con::cube in a private network, i.e. in a LAN unless the firewall is configured to forward 

external requests to the con::cube (port-forwarding).

con::cube D-330, 01-2020 Release

www.s-can.at

Copyright © s::can Messtechnik GmbH

19 / 42

!

!

Содержание con::cube D-330

Страница 1: ...Manual con cube D 330 January 2020 Release S 45 M...

Страница 2: ......

Страница 3: ...Mode 16 6 Operation of con cube 17 6 1 Direct Operation of con cube 17 6 2 Remote Operation of con cube 17 6 2 1 Remote Operation of con cube via Web Browser 17 6 2 2 Remote Operation of con cube via...

Страница 4: ...Cleaning 35 11 2 1 Cleaning Valve 35 11 3 Spare Parts 35 11 3 1 Desiccant Package 35 11 3 2 Touch Pen 35 11 4 Optional Features 36 11 4 1 Gateway to 3G Modem 36 11 4 2 Analog Output Module 36 11 4 3 D...

Страница 5: ...d in English and German by s can This original manual serves as the reference in case discrepancies occur in versions of the manual after translation into third languages This manual and all informati...

Страница 6: ...products are leaving our factory in immaculate technical and safety conditions Inappropriate or not intended use of the product however can cause danger The manufacturer is not responsible for damage...

Страница 7: ...ia analog input optional Integration of external sensor signals via digital input optional Network connectivity via ethernet WLAN or optional 3G 4G modem Remote control of s can monitoring station via...

Страница 8: ...on of third party readings input module D 315 IN RELAY 2 digital inputs for integration of third party readings input module D 315 ANTENNA PLUG internal antenna adapter cable and connector D 315 ANTEN...

Страница 9: ...le glands M16 optional for Profibus 3 cable glands M16 1 cable gland M16 optional antenna plug 2 cable glands M20 Holder for housing cover Desiccant package Insert for SIM card Additional cable termin...

Страница 10: ...s well as electromagnetic radiation Damage to the device caused by wrong storage will not be covered by warranty Transport should be done in a packaging that protects the device original packaging or...

Страница 11: ...an option Cable for power supply part no C 31 xx Gateway to 3G part no D 315 3GLX Internal antenna adapter cable and connector part no D 315 ANTENNA PLUG External antenna part no D 315 ANTENNA PRO Di...

Страница 12: ...Best possible weather and splash water proof set up Infrastructure energy data and compressed air Power supply for controller for operation operational reliability voltage power peak free Oil and part...

Страница 13: ...1 Plan of Terminal Connections Once the housing cover has been opened four hexagonal head screws removed you have access to the cable terminal compartment of the con cube Opening of the cable terminal...

Страница 14: ...ox B 23 xxx has to be used to connect a further spectrometer probe 5 3 Connection of s can ISE Probes and Sensors Alternative or additional to the spectrometer probe it is possible to connect ISE prob...

Страница 15: ...arameterised within the moni tool software please refer to manual moni tool 5 5 Connection of Main Power Supply This type of work must be performed by authorised persons only see section 2 Depending o...

Страница 16: ...display Please refer to the manual of the operation software moni tool for all further tasks to complete the initialisation of your s can monitoring system 5 7 con cube Sleep Mode For operation with l...

Страница 17: ...6 1 Direct Operation of con cube Normally this direct operation is performed by a person standing in front of the con cube and using the touch screen Alternatively moni tool can also be operated with...

Страница 18: ...once the remote connection was successful The password to be entered is scan case sensitive If you operate the con cube via VNC Viewer all actions can be seen on the touch screen of the con cube also...

Страница 19: ...contains all types of possible connectivity setups Regarding the individual connection types please refer to the subsections afterwards con cube Type of data transfer Connection type Customer infrast...

Страница 20: ...cube 192 168 10 12 IP address of the PC 192 168 10 2 Subnet mask for both devices 255 255 255 0 Connect your network cable RJ45 to the con cube and to the PC when working via LAN see section 6 5 1 Al...

Страница 21: ...ofile e g by editing the existing Default profile via Service Terminal Network Settings Modem Select Default and push icon Edit Enter a Profile name and the PIN of the SIM card Enter the Apn and the A...

Страница 22: ...ich is always OpenVPN Custom Push the button Save Within the next window enter a Profile name Select the file containing the Private key password e g password txt Select the file containing the OpenVP...

Страница 23: ...moni tool menu item Service Outputs TML Transfer of measurement result files automatical via FTP see manual moni tool Service Outputs File Output FTP Transfer automatical via SSH see manual moni tool...

Страница 24: ...eger 0xXXYY int16 has the following mapping M o d b u s r e g i s t e r Register address b15 b14 b13 b12 b11 b10 b9 b8 b7 b6 b5 b4 b3 b2 b1 b0 0xXXXX 0xXX 0xYY For example 17289 0xBC77 has the mapping...

Страница 25: ...ion Function code Description used for Read Input Register 0x04 Parameter readings Parameter status Device status Read Holding Register 0x03 Device and parameter configuration settings Mirrored input...

Страница 26: ...s 0x00C8 bitmask 1 R Para 10 status 1 200 xP10Value 0x00CA float 2 R Para 10 result 202 parameter 11 result bmP11Status 0x00D0 bitmask 1 R Para 11 status 1 208 xP11Value 0x00D2 float 2 R Para 11 resul...

Страница 27: ...0x0162 float 2 R Para 29result 354 parameter 30 result bmP30Status 0x0168 bitmask 1 R Para 30 status 1 360 xP30Value 0x016A float 2 R Para 30 result 362 parameter 31 result bmP31Status 0x0170 bitmask...

Страница 28: ...not supported or reserved Each bit reperesents a condition 1 error 0 o k M a p p i n g o f P a r a m e t e r S t a t u s b m P x S t a t u s Bit Display xPxValue Status Description b0 NaN or value Ge...

Страница 29: ...atus Status of parameter 7 0x2A 0x2D Float xP7Value Reading of parameter 7 0x2E 0x2F Bitmask bmP8Status Status of parameter 8 0x30 0x33 Float xP8Value Reading of parameter 8 0x34 0x40 TAI64n tSampleTi...

Страница 30: ...remind that polling of readings needs several seconds Automatic measurement active Readings are actualized Service mode not active Check measuring settings Leave Service mode System status LED is blu...

Страница 31: ...have to be checked for cleanliness possible damage and dirt or foreign bodies before closing the housing cover every time In case of damage to the cord gasket 2 in the housing cover it has to be repai...

Страница 32: ...rn Material Authorization Return consignments of the s can monitoring system or parts of the system shall be done in a packaging that protects the device original packaging or protective covering if p...

Страница 33: ...2 x M8 U bolt with screw nut and spring washer 4 x M5 flange nut included in delivery for mounting onto rails 11 1 2 Power Supply Cord For connection of the device to the power supply a power supply...

Страница 34: ...install antenna under large overhanging tree branches The mounting of the antenna is performed by the following steps Fix carrier brackets on mounting console with bolts Fix mounting console antenna...

Страница 35: ...2 coupler screw connection 6 4 mm nickel plated brass P inlet pipe compressor compressed air side Connection fitting probe side B 44 Hose ID 3 8 inch B 44 2 plug in fitting 6 mm OD A inlet pipe probe...

Страница 36: ...see sec tion 5 1 for correct wiring Configuration via moni tool software see section 12 1 for address mapping Assembling only by qualified personal authorized by s can con cube can be equipped with ma...

Страница 37: ...pped wire via terminal block see sec tion 5 1 for correct wiring Configuration via moni tool software see section 12 1 for address mapping Assembling only by qualified personal authorized by s can con...

Страница 38: ...Hz wide range power supply unit Power supply cable outside diameter 5 12 mm single cable wires 1 5 2 5 mm2 min 80 C min 300 VAC Required galvanic isolation of power input 3 0 kV 1 5 kV D 330 230 D 33...

Страница 39: ...ax 8 parameters Analog OUT 4 to 20 mA active Resolution 12bit 5 A Accuracy 30 A Max load 500 Ohm Galvanic isolation 1 kV D 315 OUT PROFIBUS blue module address 1 see section 11 5 4 D 315 OUT SDI12 see...

Страница 40: ...ot to be used Slot 2 51 and 52 19 and 20 35 and 36 Slot 3 53 and 54 21 and 22 37 and 38 Slot 4 55 and 56 23 and 24 39 and 40 Slot 5 57 and 58 25 and 26 41 and 42 Slot 6 59 and 60 27 and 28 43 and 44 S...

Страница 41: ......

Страница 42: ...scan Messtechnik GmbH Brigittagasse 22 24 1200 Vienna Austria Tel 43 0 1 219 73 93 0 Fax 43 0 1 219 73 93 12 office s can at www s can at...

Отзывы: