4 Installation
4.1 Environment
The con::cube is designed according to environmental protection rating IP 65 and is resistant against environment
effects. Indoor use is preferred. In case of outdoor use the installation in a cabinet or on the separately available
weather shield is highly recommended to protect the device (housing cover has to be securely closed).
Enclosure class IP 65 protection is only guaranteed if the housing cover is fixed tightly with the four screws, an operative
desiccant package is mounted inside the housing and the sealing of the housing cover is undamaged and placed correctly. In
addition all cable glands have to be sealed correctly in that way they are closed tightly with the appropriate cables or dummy
caps. All connectors must be covered with corresponding caps when not in use. Any damage caused by intrusion of water will
not be covered by the warranty.
The correct installation of measuring instruments is an important prerequisite for satisfactory operation. Therefore the following
checklist for the installation can be used to ensure that all sources for potential operational problems can be ruled out to the
greatest possible extent during the installation, allowing the s::can monitoring system to operate properly.
Installation site:
Easy accessibility (mounting, sampling, functional check, demounting)
Availability of sufficient space (probe / sensor, installation fitting, controller for operation, etc.)
Adherence to limit values (see technical specifications located at the end of this manual)
Best possible weather and splash water proof set-up
Infrastructure (energy, data and compressed air):
Power supply for controller for operation (operational reliability, voltage, power, peak free)
Oil- and particle free compressed-air supply (optional for automatic probe / sensor cleaning)
Shortest possible distances between system components (probe – controller – compressed-air supply – energy supply)
Correct dimensioning, mounting and protection of all cables and lines (non-buckling, no risk of stumbling, no damage etc.)
4.2 Mounting
For mounting and electrical installation the following tools and materials are necessary:
Hexagonal wrench key (size 5) to open housing cover
Torx wrench key (size TX 25) for fixing screws of mounting rails
2 s::can mounting rails (included in delivery)
4 screws (M5) and 4 washers for fixing the two mounting rails onto the con::cube (included
in delivery)
4 screws for fastening the con::cube on the wall (if required)
Weather shield (F-51, if required)
Power supply cable (C-31-xx, if required)
Stripping tool for power supply line
Cable end sleeves and crimper
The con::cube can be mounted quickly and easily onto a flat wall using one of the following
methodes:
With the two mounting rails [1] included in delivery. Fasten the two mounting rails onto
the backside of the con::cube using the four screws [2] and the four spring washers [3] as
shown on the right.
With the four threaded holes on the backside of the con::cube to mount the device directly
from the backside (M5 screws, not included in delivery).
For the correct dimensioning and space required for mounting please refer to the figures in
section 3.3 and the technical specifications.
con::cube D-330, 01-2020 Release
Copyright © s::can Messtechnik GmbH
www.s-can.at
12 / 42