F Ü R F U S S U N D P F L E G E
F Ü R F U S S U N D P F L E G E
Operating Instructions PODOLOG STELLA 3s | Revision 00
Use of this equipment adjacent to or stacked with other
equipment should be avoided because it could result in improper operation.
If such use is necessary, this equipment and the other equipment should be
observed to verify that they are operating normally."
Use of accessories, transducers and cables other than those specifi ed
or provided by the manufacturer of this equipment could result in increa-
sed electromagnetic emissions or decreased electromagnetic immunity
of this equipment and result in improper operation.
Guidelines and manufacturer’s declaration – electromagnetic interference immunity
The PODOLOG STELLA 3s is designed for use in electromagnetic environments as specifi ed
below. The customer or user of the PODOLOG STELLA 3s should ensure that the device is
used in the prescribed environment.
Interference immu-
nity test
IEC 60601 test level
Conformance level Electromagnetic en-
vironment guidelines
Discharging
of static
electricity (ESD)
accord-ing to IEC
61000-4-2
Contact Discharge:
±8 kV
Air Discharge:
±2 kV, ±4 kV, ±8 kV,
±15 kV
Contact Discharge:
±8 kV
Air Discharge:
±2 kV, ±4 kV, ±8
kV, ±15 kV
Floors should be made
of either wood, concre-
te or with a ceramic tile
covering. If the fl oor is
covered with a synthe-
tic material, relative air
hu-midity must be at
least 30%.
38
I
ELECTROMAGNETIC COMPATIBILITY