Operating Instructions PODOLOG STELLA 3s | Revision 00
Guidelines and manufacter’s declaration – electromagnetic interference immunity
Voltage dips,
short-break power
supply and fl uctua-
tions in the power
supply voltage
ac-cording to
IEC 61000-4-11
0 % UT for 0.5
cycle at 8 phase
angles
0 % UT for 1 cycle
at 0°
70 % UT for 25/30
cycles at 0°
0 % UT for
250/300 cycles
at 0°
0 % UT for 0.5
cycle at 8 phase
angles
0 % UT for 1 cycle
at 0°
70 % UT for 25/30
cycles at 0°
0 % UT for
250/300 cycles
at 0°
The quality of the supply
current shall correspond to
typical commercial or hospital
environment standards.
Should the user of PODOLOG
STELLA 3s require continued
functionality following a power
outage, we rec-ommend sup-
plying the PODOLOG STELLA
3s with current from an uninter-
ruptible power system or via
battery.
NOTE: UT is the supply voltage before implementation of the test level
Rated power
frequency magnetic
fi elds
(IEC 61000-4-8)
30 A/m, 50 Hz and
60 Hz
30 A/m, 50 Hz and
60 Hz
Power frequency mag-
netic fi elds should be
at level characteristics
of a typical location in
a typical commercial or
hospital enviroment.