background image

 

 

 

 

10 

Warning:

 Should it be determined that the air detection calibration failure has been caused by the presence of oxygen 

depleting compounds in the surrounding atmosphere the use MUST take immediate action to rid the area of 
such compounds or provide adequate ventilation to prevent such accumulation. The accumulation of 
oxygen depleting compounds in the surrounding atmosphere will present a human health risk and hazard. 

3.

 

To clear the error code power down the instrument. The error code will automatically clear. 

RECALIBRATION REQUIRED: 

Signifies that the prior calibration data is too old and requires refreshing. This message 

will appear whenever a 3 minute period elapses from the prior calibration. 

Corrective Actions: 

1. Remove the sample hose connection from the refrigerant source if so connected. Depress the A button to initiate 
a new calibration.  

B. Maintenance Procedures 

1.

 

Sample Filter Replacement 

Inspection Frequency:

 Inspect the sample filter before and after each use of the instrument. 

What to look for:

 Inspect the outside diameter of the white element for the appearance of any red spots or the presence 

of discoloration. The appearance of red spots or the presence of discoloration will indicate that the sample filter requires 
replacement. Do not look into the ends of the filter element since the red indicating dye may always be seen from this 
angle. 

Failure to replace the sample filter when so indicated may result in out of warranty damage to the instrument.

 

Replacement Procedure:

 

a)

 

Power down the instrument and disconnect the sample hose. 

b)

 

Obtain a replacement filter. 

c)

 

Remove the existing filter by pulling it straight up and out of the retaining clip and disconnecting the tube 
connections. Do not allow the free tube ends to slip back into the instrument. 

d)

 

Discard the existing filter. 

e)

 

Install the free ends of the instrument tubes onto the ends of the replacement filter being sure to align the filter flow 
arrow with the arrow on the instrument case. 

f)

 

Carefully slide the tubes back into the instrument case and position the filter into its retaining clip on the instrument 
case. 

g)

 

Inspect the sample hoses for signs of oil entrapment. The need to replace the sample filter may indicate sample 
hoses contaminated with oil. Replace or flush the sample hose(s) if oil entrapment is found. 

2.

 

Sample Hose Inspection and Cleaning 

Inspection Frequency:

 Inspect the sample hose(s) before and after each use of the instrument and after every 

replacement of the sample filter. 

What to look for:

 Inspect the inside diameter of the inner tube for signs of cracking, fouling, oil entrapment, kinks or 

other signs of wear. Oil contamination can by cleansed by flushing the hose(s) as described below. Hoses that show 
signs of wear should replaced immediately. 

Sample Hose Flushing Procedure:

 

a)

 

Remove the hose from the instrument and flush with isopropyl alcohol. 

Warning:

 Flush hoses away from sparks, open flames or any other ignition source and in an area with appropriate 

ventilation. 

b)

 

Dry the hose by blowing clean, dry oil-free nitrogen or shop air through the inner tube; or; allow the hose to air-
dry for several hours. Take care not to dry the hose with lubricated shop air. 

Содержание ID900

Страница 1: ...l rti rtitech com Web site www rtitech com OPERATION MANUAL ID900 Portable Refrigerant Identifier Manual No 035 80979 00 WARNING Liquid Samples will Damage Unit Do not operate if filter has red flecks...

Страница 2: ...1 Power Connector 5 B Sample Hoses 6 1 R12 Sample Hose 6 2 R134a Sample Hose 6 C Power Harnesses 6 1 Vehicle Battery Adapter 6 2 Wall Power Adapter 6 D R134a Tank Adapter Fitting 6 E Storage Case 6 IV...

Страница 3: ...e Input Warning The instrument requires connection of the sample hose to the low or vapor side port of the source vehicle or refrigerant cylinder Connection of the sample hose to the high or liquid po...

Страница 4: ...a storage case that will house all components R12 and R134a sample hoses are provided as well as a R134a cylinder adapter fitting and a power cord for connection directly to the vehicle battery An op...

Страница 5: ...as been analyzed by the instrument 7 Sample Inlet Port permits the connection of either a R12 or R134a sample hose to the instrument 8 Air Detection Sensor a chemical fuel cell that will detect the am...

Страница 6: ...end This harness is utilized to connect power from the vehicle battery to operate the instrument and avoid the use of extension cords 2 Wall Power Adapter Optional is a 6 foot power cable that has an...

Страница 7: ...ect the performance of the instrument Once the elevation has been entered it does not ever require entry again unless the instrument is moved to a new elevation The screen USAGE ELEVATION NOT SET is d...

Страница 8: ...or Hydrocarbons have been detected in concentrations of 4 by weight or greater No weight concentrations of refrigerants or air are provided in this mode d NO REFRIGERANT CHK HOSE CONN Indicated whene...

Страница 9: ...ensor failure due to temperature extremes Corrective Actions 1 Allow the instrument to stabilize at room temperature for at least 30 minutes before attempting use 2 To clear the error code power down...

Страница 10: ...Procedure a Power down the instrument and disconnect the sample hose b Obtain a replacement filter c Remove the existing filter by pulling it straight up and out of the retaining clip and disconnecti...

Страница 11: ...from the instrument off of the support pins and remove the sensor by unthreading Discard the spent sensor g Obtain a replacement air detection sensor part number 1 16 1013 00 0 and install it into the...

Страница 12: ...erant R12 or R134a is detected in concentrations equal to or greater than 98 by weight Unknown Refrigerant Purity Setpoint In the presence of detected unknown refrigerant types R12 or R134a is conside...

Отзывы: