MEMS Tilt Beam Instruction Manual
RST Instruments
8
3.4 E
LECTRICAL
C
ONNECTIONS
1. Under the Colour Code Table, look up the lead designations for the type of
cable being used.
2. Under the Electrical Connections Table, make the appropriate lead connections,
according to the type of system being used.
3. Horizontal and Vertical Tilt Beam system measure the A-axis (Green & White).
Vertical Tile Beam system is capable of measuring the B-axis (optional) (Orange
& Blue).
Information regarding your sensor configuration and cable type is listed below and on
your Calibration Certificate.
Figure 7: Electrical Connections
Tiltmeter
Wire Color
Terminal
Datalogger
12V
Red
1
12V
Gnd
Black
2
Gnd
A+
Green
3
xH
A-
White
4
xL
B+
Orange
5
yH
B-
Blue
6
yL
Note: For
BUSSED
Systems only, use the cable with the jacket stripped 25mm back
for the Cable Output.