1. Emissivity:
Emissivity is a value between 0.17 and 1 which
indicates an object’s ability to emit infrared energy. Emissivity
is determined primarily by the objects’ composition and surface
finish.
The thermometers’ sensitivity to emissivity “
ε
” is preset at
0.95, because this setting will be applicable to 90% of
surfaces, but can be adjusted as required.
2. Emissivity adjustment
c
Apply black tape, matt-black paint or black magic marker to
the object if it is safe to do so.
d
“
ε
” is set at 0.95 to indicate correctly from a dark surface.
e
Aim the sensor at the black surface, pull the trigger and
note the reading as T1.
f
Remove the black tape or matt-black paint and aim the
sensor at the same area again. Pull the trigger and note the
reading as (T2).
g
Change the emissivity value “
ε
” until T2 and T1 are equal
as follows:
a). Enter setting mode (refer to
MODE
button above).
b). Press
MODE
button to select the
ε
value setting.
c). Press
▲
or
▼
buttons to set desired value.