SPECIFICATIONS / EN
EN-2
10/15/10 Version No. 00
2. SPECIFICATIONS
2-1 Electrical specifications
Measurement range:
J - type: -150.0 to +1090.0°C (-200.0 to +1994.0°F)
K - type: -150.0 to +1370.0°C (-200.0 to +1999.9°F)
T - type: -150.0 to +400.0°C (-200.0 to +752.0°F)
E - type: -150.0 to +870.0°C (-200.0 to +1598.0°F)
N - type: -150.0 to +1300.0°C (-200.0 to +1999.9°F)
R - type: 2.0 to +1767.0°C (+35 to +1999.9°F)
S - type: 2.0 to +1767.0°C (+35 to +1999.9°F)
Display effective resolution:
J, K, T, E and N type: 0.1°C /°F /K
R and S type: 1.0°C
/
°F /K (0.1°C
/
°F /K only for reference)
Measurement accuracy:
J, K, T, E and N type:
±
[0.05% of r0.5°C (0.9°F)]
[
Below -100°C (-148°F): add 0.15% of reading for J, K, E
and N; and 0.45% of reading for T
]
R and S type:
±
[0.05% of r2°C (4°F)]
NOTE
This basic accuracy specification does not include the
error of the temperature probe. Please refer to the
temperature probe accuracy specification for additional
details.