16
Galaxy S+ CO
2
Incubator
User Guide
3
3
O
O
P
P
E
E
R
R
A
A
T
T
I
I
O
O
N
N
3.1
Control Panel
The control panel consists of two individual three-digit LED displays, and four
Function keys (
see Figure 9 below
).
Figure 9: Control Panel
Table 3: Function Keys
Key Symbol
Key Name
Function
∗
Programming
Press this key to enter Programming mode, and to
set values in either display.
-
Down
In Programming mode, use this key to scroll down
through numbered values in the display. Press this
key simultaneously with the Enter key to access
the Alarm system.
+
Up
In Programming mode, use this key to scroll up
through numbered values in the display.
↵
Enter
Press this key to save a new set point. Press this
key simultaneously with the Down key to access
the Alarm system.
#
NOTE:
If you accidentally press both the Up and Down keys simultaneously, you will
engage Engineering Mode: press the Programming key immediately to exit.
CO
2
DISPLAY
(shown in normal
operation at factory
setpoint)
TEMPERATURE
DISPLAY
(shown in normal
operation at factory
setpoint)
FUNCTION KEYS