Roth Cyclo II Скачать руководство пользователя страница 6

gk  02/2006 

6

 

ƒ

  In exceptional cases, too loose insertion of the tube – see paragraph 5. (Insert and replace tube) on 

changing tube – can also lead to slight jumping at the discharge side clamping jaws. In this case, it is 
sufficient to straighten the tube by pulling slightly and increase the pressure somewhat. 

 

Malfunction Possible 

cause 

Remedy 

Pump  
not running 

  Pump not plugged into mains 

 Fuse 

defective 

  Rotor stuck because wrongly assembled 

 Insert 

plug 

 Change 

fuse 

19

 

  Assemble pumping head according 

to instructions in 9. 

Reduced pumped quantities, 
tube is crushed 

  Wrong tube size 

  Use tube suitable for tube cartridge 

Pump does not pump evenly 

  Tube material used unsuitable for the 

medium  

 Rollers 

soiled 

  Wrong tube size 

  Use suitable tube (see list) 

  Dismantle head, clean thoroughly  

  Use tube suitable for tube cartridge 

 

If the pump is shut down, especially for several hours or days, the tube should be relieved of the pressure, 
otherwise a permanent deformation of the tube can be induced and the pump will no longer work perfectly. 
 
 

11. Technical Data 

 

Dimensions [WxHxD] 

134 x 131 x 190 mm 

Weight 1.700 

Voltage 230 

VAC 

Fuse 

T 0.2 A / 250V 

Drive DC-Motor 

with 

planetary 

gears 

Power delivery 

10 W 

 

Flow rate  

Tube ID x Thickness 

ml/min (3-roll head) 

ml/min (6-roll head) 

1.6 x 1.6  

1.5 - 30 

1.2 - 24 

3.2 x 1.6  

6 - 110 

4.4 - 88 

4.8 x 1.6  

12 - 240 

9.6 - 192 

 
 

Permissible pressure 1 bar ! In special cases (only after release) max. 3 bar. The tube manufacturer’s 
directions must be observed. 

 
We retain the right to make technical amendments! 
 

In case of unauthorised opening, repair or alteration to the appliance, all guarantee and liability 
claims are lost 

 

 
 

12. Accessories: 

 

Tube cartridges: 

For tube inner Ø x 
thickness mm  

Tube cartridges  
Art. No. 

1.6 x 1.6  

Y375.1 

3.2 x 1.6 

Y376.1 

 

 

Содержание Cyclo II

Страница 1: ...ence This instruction manual will familiarize you on how to operate and use our pericyclic pump Roth Cyclo ll Please read it thoroughly before operating the unit Operating Instructions Page 1 Delivery includes 2 2 Safety instructions 2 3 Areas of Application 2 4 Initial operation 3 5 Tube cartridges and Tubes 3 5 1 Selection of the correct tube cartridge 3 5 2 Inserting and replacing Tube 4 6 Sett...

Страница 2: ...ched off mains plug removed Only authorised original spare parts accessories may be used The use of other parts conceals unknown risks and must be refrained from in every case Any necessary checks service or repair work may only be carried out by Carl Roth GmbH Co KG otherwise we are unable to guarantee the operatability and safety of the unit Carl Roth GmbH Co KG is not liable for damage to the m...

Страница 3: ...e a tube is inserted and the tube changed it must be ensured that the cartridge is designed for the tube used Only pump hoses with a wall thickness of 1 6 mm are suitable for use Hose ID mm hose cassette 1 6 1 6 Y375 1 3 2 3 2 Y376 1 4 8 4 8 included in delivery 3 roll head Pumping capacity ml min 1 5 30 6 110 12 240 6 roll head Pumping capacity ml min 1 2 24 4 4 88 9 6 192 Fig 3 Individual parts ...

Страница 4: ... ensure pumping reliability 6 Secure lock nuts 2 by turning to the right Important If counter pressure exists the pressure setting must be carried out under operating pressure This aforementioned setting can now be retained unchanged for the same operating conditions i e for the same counter pressure and the same pumping tube Changing the tube is therefore significantly simplified as only the pivo...

Страница 5: ...ody can be removed by loosening the 3 crosshead screws and then cleaned Rinse off all removable parts with tap water if necessary with cleaning additives dry off and reassemble in the reverse order Attention Place the roller rotor 5 on the drive shaft so that the driving pin 13 sits in the groove on the rear of the roller rotor 5 Occasional cleaning as described above is urgently recommended after...

Страница 6: ...roughly Use tube suitable for tube cartridge If the pump is shut down especially for several hours or days the tube should be relieved of the pressure otherwise a permanent deformation of the tube can be induced and the pump will no longer work perfectly 11 Technical Data Dimensions WxHxD 134 x 131 x 190 mm Weight 1 700 g Voltage 230 VAC Fuse T 0 2 A 250V Drive DC Motor with planetary gears Power ...

Отзывы: