gk 02/2006
6
In exceptional cases, too loose insertion of the tube – see paragraph 5. (Insert and replace tube) on
changing tube – can also lead to slight jumping at the discharge side clamping jaws. In this case, it is
sufficient to straighten the tube by pulling slightly and increase the pressure somewhat.
Malfunction Possible
cause
Remedy
Pump
not running
•
Pump not plugged into mains
•
Fuse
defective
•
Rotor stuck because wrongly assembled
•
Insert
plug
•
Change
fuse
19
•
Assemble pumping head according
to instructions in 9.
Reduced pumped quantities,
tube is crushed
•
Wrong tube size
•
Use tube suitable for tube cartridge
Pump does not pump evenly
•
Tube material used unsuitable for the
medium
•
Rollers
soiled
•
Wrong tube size
•
Use suitable tube (see list)
•
Dismantle head, clean thoroughly
•
Use tube suitable for tube cartridge
If the pump is shut down, especially for several hours or days, the tube should be relieved of the pressure,
otherwise a permanent deformation of the tube can be induced and the pump will no longer work perfectly.
11. Technical Data
Dimensions [WxHxD]
134 x 131 x 190 mm
Weight 1.700
g
Voltage 230
VAC
Fuse
T 0.2 A / 250V
Drive DC-Motor
with
planetary
gears
Power delivery
10 W
Flow rate
Tube ID x Thickness
ml/min (3-roll head)
ml/min (6-roll head)
1.6 x 1.6
1.5 - 30
1.2 - 24
3.2 x 1.6
6 - 110
4.4 - 88
4.8 x 1.6
12 - 240
9.6 - 192
Permissible pressure 1 bar ! In special cases (only after release) max. 3 bar. The tube manufacturer’s
directions must be observed.
We retain the right to make technical amendments!
In case of unauthorised opening, repair or alteration to the appliance, all guarantee and liability
claims are lost
12. Accessories:
Tube cartridges:
For tube inner Ø x
thickness mm
Tube cartridges
Art. No.
1.6 x 1.6
Y375.1
3.2 x 1.6
Y376.1