CONTROL DIAL SETTINGS
(continued)
8
DIAL
FUNCTION
HOLD
Stops pumping action and audio alarms.
HOLD
is used when correcting an alarm
condition or whenever pumping action is
temporarily interrupted (such as when con-
necting a new feeding container). If pump
is left on
HOLD
for more than 5 minutes,
an alarm will sound and the message
SEL
run
(select RUN) will be displayed.
NOTE:
The
HOLD
timer may be pro-
grammed for a duration of 1-90 minutes (in
1-minute intervals) by depressing the key-
pad arrows either up or down when the
dial is in the
HOLD
position.
OFF/
CHARGE
SET RATE
ml/hr
SET DOSE
ml
RUN
ml/hr
HOLD
VOL FED
ml
CLEAR
VOL
H 5
OFF/
CHARGE
SET RATE
ml/hr
SET DOSE
ml
RUN
ml/hr
HOLD
VOL FED
ml
CLEAR
VOL
980
OFF/
CHARGE
SET RATE
ml/hr
SET DOSE
ml
RUN
ml/hr
HOLD
VOL FED
ml
CLEAR
VOL
0000
CLEAR VOL
Stops pumping action and resets the volume
fed to zero.
NOTE:
The pump will “beep” four times
before clearing the volume to alert users
that the dial has been turned to
CLEAR
VOL
. The user must turn dial away from
CLEAR VOL
(to any other setting) before
the fourth beep to avoid an unintentional
loss of accumulated volume.
VOL FED
Stops pumping action. The display will
show the volume delivered. Pump accumu-
lates volume delivered up to 9999 mL
unless the volume is reset to zero by turning
to
CLEAR VOL
(or the pump is unplugged
while in the
OFF/CHARGE
mode). Upon
exceeding the maximum volume allowed
on the display, the value turns to zero and a
new accumulation begins. This is not
accompanied by an alarm.