CSA Standard C22.2 No. 125
17
SPECIFICATIONS
Power
Voltage:
100 to 130 VAC, 60 Hz, 1-Phase
Power:
15 watts, average
Fuse:
200 mA, time-delay type
Line Cord:
Hospital-grade, nondetachable (10 feet)
Leakage:
Less than 100 microamps
Mechanical
Height:
8.5 inches
Width:
6.5 inches
Depth:
4.8 inches
Weight:
6.6 lbs
Operational Specifications — Flow Rates
Range:
1-300 mL/hr
Increments:
1 mL/hr
Accuracy:
±10% or ±0.5 mL/hr, whichever is greater (refer to
CHECKING PUMP ACCURACY, page 12)
Pressure:
Pumps against 18 psi nominal back pressure before
no FLO alarm. Refer to PATROL
®
MAINTENANCE AND REPAIR
MANUAL for test procedure.
Battery Operation
Type:
Rechargeable sealed lead acid
Voltage:
6 V; the pump will fully recharge in a maximum of 4.5 hours
while pump is off, or 12 hours if the pump is operating.
Expected Life: When fully charged, 8 hours at 125 mL/hr.
Standards
Designed and manufactured to meet requirements of
52034/52036-M-21395-PTR.a 6/14/01 3:06 PM Page 17