INSTRUCTIONS FOR USE
(continued)
11
19. When feeding is complete, turn
pump dial to
OFF
.
NOTE:
If alarm sounds, turn
pump dial to
HOLD
. Correct
alarm condition indicated by visu-
al display, then turn dial to
RUN
to restart feeding. (For explana-
tion of alarms, see pages 13 and
14.)
PRECAUTION:
The pump set
and container should be replaced
as needed, or at least every 36
hours, to avoid fluid-contamina-
tion problems. For single-patient
use only. Do not store for pro-
longed periods at extreme tem-
peratures.
18. To clear the volume delivered,
turn dial to
CLEAR VOL
. (This
occurs after four beeps.)
17. To see feeding volume delivered,
turn dial to
VOL FED
.
16. Turn dial to
RUN
to start feeding.
15. If desired, turn dial to
SET DOSE
and select dose by pressing the
arrows.
NOTE:
Pump will operate with-
out a
DOSE
selected (SET DOSE
at 0000).
14. Select flow rate, from 1 to 300
mL/hr, by pressing the arrows and
observing the entry on the dis-
play. Holding down an arrow
will cause the numbers to scroll.
Never use sharp objects such as
pens, pencils, fingernails or paper
clips to depress the keypad
arrows.
TO START PUMP:
13. Turn pump dial to
SET RATE
.
Pump will initiate system self-
check procedure.
OFF/
CHARGE
SET RATE
ml/hr
SET DOSE
ml
RUN
ml/hr
HOLD
VOL FED
ml
R
75
OFF/
CHARGE
SET RATE
ml/hr
SET DOSE
ml
RUN
ml/hr
HOLD
VOL FED
ml
CLEAR
VOL
980