June 2017
6
Quick Start Guide
5.0 Eliminate mounting effects
Devices are factory-calibrated. Once installed, it is recommended to perform this
step to eliminate potential error caused by mounting position or static pressure.
Instructions for using a Field Communicator are listed below.
Note
See the Rosemount Wireless Pressure Gauge
Reference Manual
for the following:
Using AMS Wireless Configurator
Sensor trim function on absolute gauge
1. Vent the device.
2. Connect the Field Communicator.
3. From the
HOME
screen, enter the HART Fast Key sequence.
4. Follow the commands to perform the procedure.
6.0 Activate wireless
Do not activate wireless until Smart Wireless Gateway is installed and functioning
properly; toggling off and on reduces battery life.
Note
If Network ID and Join Key is specified at order entry, then the device will automatically
search and connect to the wireless network when powered on.
6.1 Join device to network
1. Obtain Network ID and Join Key for the wireless network (available in wireless
gateway).
2. From the
HOME
screen, enter the HART Fast Key sequence.
3. Follow the commands to perform the procedure.
4. Select
Overview
>
Status
.
5. Verify the communication status displays
Connected.
Note
Joining the device to the network could take several minutes.
Device dashboard Fast Keys
2, 1, 1
Device dashboard Fast Keys
2, 1, 2