Limitations

• Current through terminals 1 and 2 shall not exceed 600mA (peak)
• Current through terminals 18 and 19 shall not exceed 5mA.
• Calibration shall occur with the standard set at 30 feet per second. 

 

Procedure

1. Set-up your equipment according to Figure 3.

2. Power up the system.

3. Allow the Rosemount 8714D to warm-up for 30 minutes.

4. Set the input current to terminals 1 and 2 at 500mA.

5. Measure and record the average voltage from terminals 18 and 19 over a five minute 

period.

6. Measure and record the average voltage across the 1 ohm resistor.

7. Calculate current with V=IR where R is the resistor value.

8. Divide result from step 5 by step 7 result.

9. Power down the system.

10. Switch the connections 1 and 2 at the Rosemount 8714D calibration standard.

11. Set the input current to terminals 1 and 2 at 500 mA.

12. Measure and record the average voltage from terminals 18 and 19 over a five minute 

period.

13. Measure and record the average voltage across the 1 ohm resistor.

14. Calculate current with V=IR where R is the resistor value.

15. Divide result from step 12 by step 14 result.

16. Add the absolute value steps 8 and 15 and divide by two. The value should be within 

0.05% (0.00108 mV/Amp) of 2.156mV/Amp.

17. If your device does not operate in conjunction with this specification, return the 

Rosemount 8714D to the Rosemount Inc. factory for adjustment.

Figure 3.  Set-up Rosemount 8714D and the transmitter as shown

Within the United States, Rosemount Inc. has two toll-free assistance numbers.

Customer Central:

 

1-800-999-9307 
(7:00 a.m. to 7:00 p.m. CST)
Technical support, quoting, and order-related questions.

North American Response Center:

1-800-654-7768 
(24 hours a day - Includes Canada) Equipment service needs.

Current 

Source

Rosemount 

8714D

Voltmeter

1 Ohm

Resistor

Current 

Source

18

19

17

1

2

4764_revBA_QIG.fm  Page 5  Tuesday, June 22, 2004  1:49 PM

Содержание 8714D

Страница 1: ...00825 0100 4764 Rev BA June 2004 Rosemount 8714D 00825 0100 4764W Start Step 1 Change Transmitter Parameters Step 2 Connect the Rosemount 8714D Step 3 Perform Electronics Trim Step 4 Verify 8714D Calibration Standard Magnetic Flowmeter Simulator ...

Страница 2: ... provides basic guidelines for the Rosemount 8714D It does not provide instructions for detailed configuration diagnostics maintenance service troubleshooting The manual and this QIG are also available electronically on www rosemount com WARNING Explosions could result in death or serious injury In an Explosion proof Flame proof installation do not remove the transmitter covers when power is appli...

Страница 3: ...etailed instructions regarding 8712D 8732C and 8742C Rosemount 8712D Figure 1 Use the six contact Pin terminal plug style wiring assembly Follow the numeric convention for the 8712D so the plugs match the terminal block Rosemount 8732C 8742C Figure 2 Use the two black connector style wiring assembly One connector has four contacts the other has three Remove the electronics side cover furthest away...

Страница 4: ...e original configuration If the reading is not within this range continue to step 8 4 Initiate an electronics trim with the LOI or HART communicator The electronics trim takes about six minutes to complete No transmitter adjustments are required STEP 4 VERIFY ROSEMOUNT 8714D CALIBRATION STANDARD Guidelines for Checking Series 8714D Magnetic Flowmeter Calibration Standard The Rosemount 8714D Magnet...

Страница 5: ...d 11 Set the input current to terminals 1 and 2 at 500 mA 12 Measure and record the average voltage from terminals 18 and 19 over a five minute period 13 Measure and record the average voltage across the 1 ohm resistor 14 Calculate current with V IR where R is the resistor value 15 Divide result from step 12 by step 14 result 16 Add the absolute value steps 8 and 15 and divide by two The value sho...

Страница 6: ...Quick Installation Guide 00825 0100 4764 Rev BA June 2004 Rosemount 8714D 6 ...

Отзывы: