Rosemount 8714D

4

CAUTION

Attempting an electronics trim without a Rosemount 8714D may result in an inaccurate transmitter. It 
may also generate a DIGITAL TRIM FAILURE message. If this message appears, no values were 
changed in the transmitter. Simply power down the transmitter to clear the message. If the trim was 
completed, or no error message was generated, correction requires a Rosemount 8714D. 

S

TEP

 3: P

ERFORM

 E

LECTRONICS

 T

RIM

1. Set the Rosemount 8714D to simulate a flow rate of 30 ft/s.
2. Power up the transmitter with the Rosemount 8714D connected. Allow the electronics 30 

minutes to warm-up before reading the flow rate.

3. Read the flow rate. It should be between 29.97 and 30.03 ft/s. If the reading is within this 

range, return the transmitter to the original configuration. If the reading is not within this 
range, continue to step 8.

4. Initiate an electronics trim with the LOI or HART communicator. The electronics trim 

takes about six minutes to complete. No transmitter adjustments are required.

S

TEP

 4: V

ERIFY

 R

OSEMOUNT

 8714D C

ALIBRATION

 S

TANDARD

Guidelines for Checking Series 
8714D Magnetic Flowmeter Calibration Standard

The Rosemount 8714D Magnetic Flowmeter Calibration Standard operates in conjunction 
with 8712D, 8732C, or 8742C Magnetic Flowmeter Transmitters. This device simulates a 
flowtube coil load. It accurately produces a simulated flow signal for the purpose of 
transmitter calibration.

Required Equipment

• Current source (output and accuracy: 500mA ±5mA)
• Precision Resistor: measured to 5 significant digits, less than 10 ppm, and > 1 watt
• Nanovoltmeter (Agilent 34420A or equivalent)

Figure 2.  Rosemount 8732C/8742C Terminal Block

HART Fast Keys

1, 5

LOI

Auxiliary Function

Coil

Connection

(Four-position

connector)

Electrode 

Connection 

(Three-position 

connector)

4764_revBA_QIG.fm  Page 4  Tuesday, June 22, 2004  1:49 PM

Содержание 8714D

Страница 1: ...00825 0100 4764 Rev BA June 2004 Rosemount 8714D 00825 0100 4764W Start Step 1 Change Transmitter Parameters Step 2 Connect the Rosemount 8714D Step 3 Perform Electronics Trim Step 4 Verify 8714D Calibration Standard Magnetic Flowmeter Simulator ...

Страница 2: ... provides basic guidelines for the Rosemount 8714D It does not provide instructions for detailed configuration diagnostics maintenance service troubleshooting The manual and this QIG are also available electronically on www rosemount com WARNING Explosions could result in death or serious injury In an Explosion proof Flame proof installation do not remove the transmitter covers when power is appli...

Страница 3: ...etailed instructions regarding 8712D 8732C and 8742C Rosemount 8712D Figure 1 Use the six contact Pin terminal plug style wiring assembly Follow the numeric convention for the 8712D so the plugs match the terminal block Rosemount 8732C 8742C Figure 2 Use the two black connector style wiring assembly One connector has four contacts the other has three Remove the electronics side cover furthest away...

Страница 4: ...e original configuration If the reading is not within this range continue to step 8 4 Initiate an electronics trim with the LOI or HART communicator The electronics trim takes about six minutes to complete No transmitter adjustments are required STEP 4 VERIFY ROSEMOUNT 8714D CALIBRATION STANDARD Guidelines for Checking Series 8714D Magnetic Flowmeter Calibration Standard The Rosemount 8714D Magnet...

Страница 5: ...d 11 Set the input current to terminals 1 and 2 at 500 mA 12 Measure and record the average voltage from terminals 18 and 19 over a five minute period 13 Measure and record the average voltage across the 1 ohm resistor 14 Calculate current with V IR where R is the resistor value 15 Divide result from step 12 by step 14 result 16 Add the absolute value steps 8 and 15 and divide by two The value sho...

Страница 6: ...Quick Installation Guide 00825 0100 4764 Rev BA June 2004 Rosemount 8714D 6 ...

Отзывы: