B-2
1.2 INSTRUMENT PHYSICAL SPECIFICATIONS
Housing
: NEMA 4X, IEC-529 IP65
Size:
1/4 DIN 43700, 92 x 92 x 156 mm (3.6 x 3.6 x 6.2
inches) cutout
Net Weight:
0.75 Kg / 1.6 lbs
Shipping Weight:
1.2 Kg / 1.6 lbs
Display:
Custom graphics liquid crystal display.
Back light: LED array defused
Resolution: 128 x 64 pixels.
1.3 ELECTRICAL SPECIFICATIONS
Power Input:
100-400VDC
,
95-240VAC ±10%, 50-60
Hz, 8 Watts max.
Conductivity:
Stability; ±0.02% /
0
C, ±0.1% 85-264VAC, 47-65Hz
Sensor Temperature Range; 0
0
C to 100
0
C (32 to 212
0
F)
Temperature Accuracy: ±0.1
0
C (0 - 100
0
C)
Repeatability: ±0.5% of span
In the Ranges listed below, the desired Span is software
selectable from the keypad by the user.
* Dependent on Liquid Type. See
pH:
Measurement Range: -4 to 16 pH (pH), -1400 to
1400 mV (ORP)
Accuracy: ±.02 pH from -4 to 16 pH, ±2 mV -1400
to 1400 mV (ORP)
Repeatability: ±.01 pH, ±1 mV (ORP)
Stability: ±.01 pH, ± 1 mV (ORP) over 30 days,
non-cumulative
Temperature Accuracy: ± 0.1
0
C
Temperature Compensation: Pt100 RTD, Automatic
glass compensation, user selectable linear solution
compensation.
RFI/EMI:
EN50081-2
EN50082-2
EN61010-1
Operating Range Specifications:
Temperature: Operating: 0
0
C to 50
0
C (32 to 122
0
F)
Storage: -20
0
C to 60
0
C (4 to 140
0
F)
Humidity: 0 to 65% RH non-condensing 0-50
0
C,
(32 to 122
0
F).
65-95% RH non-condensing 35-50
0
C
(95 to 122
0
F) 48 hour max.
1.4 ALARMS
Control Alarms:
2 SPST relays, contacts rated at 5
amps, 240 VAC with a resistive load. May be set as
normally open or normally closed. Alarms can be
configured HIGH/HIGH, HIGH/LOW, LOW/LOW, or
LOW/HIGH. Alarm deadband is adjustable over full
range.
Error Alarm:
SPST relay, used to indicate an instrument
or sensor error.
•
2.0A @ 30VDC or 60VAC ±10%.
•
1.0A @ 60VDC or 120VAC ±10%
•
0.5A @ 120VDC or 240VAC ±10%
(All resistive loads)
Range Alarm
: 1 SPST relay, used to indicate current
output range; contacts open when on Low range and
closed on High range.
•
2.0A @ 30VDC ±10%.
•
1.0A @ 60 VDC or 120VAC ±10%.
•
0.5A @ 120VAC or 240VAC ± 10%.
Outputs - Standard
Current Output :
Isolated 0/4-20mA DC into a load
resistance of 600 Ohms maximum. 600 VAC isolation.
RS485:
Standard 2 wire; 1,000 feet maximum. Non-iso-
lated.
Analog Outputs - Optional
Current Output:
Isolated 0/4-20mA DC into a load
resistance of 600 Ohms maximum. 600 VAC isolation.
Voltage Output:
Isolated 0-5 VDC into a load resis-
tance of 1,000 Ohms minimum. 600 VAC isolation.
1.5 OPTIONAL DIGITAL OUTPUTS
RS485:
Standard 2 wire; 1,000 feet maximum. 600 VAC
isolation.
RS232E:
Standard 3 wire; 1,000 feet maximum. 600
VAC isolation.
Range
Cell
Accuracy
(µS/cm)
Constant
(% of Reading)
0-50
0.01/cm
±1% or 0.002
µ
S/cm, whichever number is larger
0-10,000
1.0/cm
±2% or 2.0
µ
S/cm, whichever number is larger
500-500,00*
0.85/cm
±2% or 0.5 mS/cm, whichever number is larger
(4 electrode)
Dual Alarms
Dual Alarms are standard on the Solu Cube. The set-
point for each alarm is independently adjustable over
the full range of measurement. The relays operate either
as high or low alarms and may be set as either normally
open or normally closed. The alarm conditions are dis-
played on the analyzer as Alarm 1 and Alarm 2 and
may be configured as HIGH/LOW, HIGH/HIGH.
LOW/HIGH, or LOW/LOW.
In addition, two low level alarms are available. One may
be used for Autoranging (see standard Output) and the
other as an error alarm.
MODEL 2700
APPENDIX B
SALES AND PART ORDERING INFORMATION