DISPLAY MESSAGE
DESCRIPTION
TYPE
BAROMETER
System Barometer
WARNING
CASE TEMP
Case Temperature
WARNING
CHOP SPEED
Chopper Speed
WARNING
CRUDE NOISE
Calculated Noise
WARNING
DET SIG
Detector Signal
WARNING
DET TEMP
Detector Temperature
WARNING
LIN ERROR
Linearizer Error
WARNING
N15 VOLTS
Power Supply, -15V
WARNING
P12 VOLTS
Power Supply, +12V
WARNING
P15 VOLTS
Power Supply, +15V
WARNING
P24 VOLTS
Power Supply, +24V
WARNING
P5 VOLTS
Power Supply, +5V
WARNING
PERCENT MOD
Percent Modulation
WARNING
RAW SIGNAL
Raw Signal
WARNING
SVFLOW
Sample Bypass Flow
WARNING
SW ERROR
Software Error
FAILURE
2. Using the
↓
arrow, select Zero Ranges, press ENTER and, using the up/down
arrows, toggle to SEPARATE. Do the same for the Calibrate Ranges selection. Do
not press ESCAPE at any time unless retention of prior settings is desired.
3. Return to the
Main Menu
and make the following selections: Expert Controls and
Setup, Expert Controls, CAL DATA softkey, FACTORS softkey, and Range 1 (2, 3,
4) Factors (do Steps 4 and 5 for each range).
4. Select Zero Offset, press ENTER, adjust the value to 500000 with the
↑
and
↓
arrow keys, and press ENTER. Do not press ESCAPE at any time unless retention
of prior settings is desired.
5. Refer to the Data Sheet in the rear of this manual for Span Factors as originally set
at the factory. Select Span Factor, press ENTER, adjust the value to match the
values on the Data Sheet with the
↑
and
↓
arrow keys, and press ENTER. If Data
Sheet is not available, enter 0.000015 with the
↑
and
↓
arrow keys, and press
ENTER. Do not press ESCAPE unless retention of prior settings is desired.
6. Attempt to recalibrate the Analyzer Module according to the procedure outlined at
the beginning of Section 3.4. If recalibration fails, return to the
Range Factors
menu, readjust factors and try calibrating again.
Another cause of failure to calibrate is the following: The value for "Maximum range"
is lower than the upper limit value for the range in use. See the
Range Settings
menu for this information (See Figure 3-11).
T
ABLE
3-1. NDIR
A
NALYZER
M
ODULE
A
LARMS
Содержание NGA 2000
Страница 1: ...NGA 2000 NON DISPERSIVE INFRARED ANALYZER MODULE Rosemount Analytical ...
Страница 6: ...CONTENTS iv June 1997 Rosemount Analytical 748332 D NGA Non Dispersive Infrared Analyzer NOTES ...
Страница 16: ...PREFACE P 10 June 1997 Rosemount Analytical 748332 D NGA Non Dispersive Infrared Analyzer NOTES ...
Страница 22: ...INTRODUCTION 1 6 June 1997 Rosemount Analytical 748332 D NGA Non Dispersive Infraraed Analyzer NOTES ...
Страница 54: ...REPLACEMENT PARTS 5 2 June 1997 Rosemount Analytical 748332 D NGA 2000 Non Dispersive Infrared Analyzer NOTES ...
Страница 58: ... blank ...
Страница 60: ... blank ...
Страница 62: ... blank ...