3.4 BINDING
To achieve full coordination between Analyzer Modules and associated I/O Modules,
the user must bind those components together in the System Set Up portion of the
Technical Configuration Menu
in software. (See Figure 3-12 of this manual and
Section 1.5 of the I/O Modules manual for binding instructions.)
3.5 CALIBRATION
Calibration can be executed from the
Basic Controls
menu. Calibration gas data can
be entered only through
the Expert Controls and Setup
menu. See Figures 3-9 and
3-11 for display screen paths.
To calibrate the Analyzer Module, introduce zero gas into the SAMPLE INLET, and do
the following:
1. If the multi-Analyzer Module, split Run Mode display is shown, press the DISPLAY
softkey until the desired Analyzer's Run Mode display is acquired.
2. Press the MENUS softkey to enter the
Main Menu
and make the following
selections from the
Main Menu
: Expert Controls and Setup, Analyzer Module
Setup, Calibration Gases.
3. Input appropriate data in the
Calibration Gas List
menu.
4. Press the HOME softkey to return to the
Main Menu.
5. Use the
↓
arrow key to select Basic Controls.
6. Press the ZERO softkey to enter the
Analyzer Zero
menu, press ZERO again and
wait.
7. Introduce span gas into the SAMPLE INLET, press SPAN softkey to enter the
Analyzer Span
menu, press SPAN again and wait.
8. Press HOME to re-enter the
Main Menu.
9. Press DISPLAY softkey for the Run Mode display.
If the user is unable to calibrate the Analyzer Module (i.e., when ZERO or SPAN is
initiated, nothing happens), a possible solution relates to the use of an incorrect gas
for zeroing or spanning (e.g., using a high concentration gas to zero or a zero gas to
span the Analyzer Module). Simply recalibrating with the appropriate gas(es) will not
correct the problem because the ZERO OFFSET or SPAN FACTOR has been set to
an extreme value in the process.
To remedy the problem, do the following:
1. Select the following from the
Main Menu
: Expert Controls and Setup, Analyzer
Module Set Up, and Calibration Parameters.
Содержание NGA 2000
Страница 1: ...NGA 2000 NON DISPERSIVE INFRARED ANALYZER MODULE Rosemount Analytical ...
Страница 6: ...CONTENTS iv June 1997 Rosemount Analytical 748332 D NGA Non Dispersive Infrared Analyzer NOTES ...
Страница 16: ...PREFACE P 10 June 1997 Rosemount Analytical 748332 D NGA Non Dispersive Infrared Analyzer NOTES ...
Страница 22: ...INTRODUCTION 1 6 June 1997 Rosemount Analytical 748332 D NGA Non Dispersive Infraraed Analyzer NOTES ...
Страница 54: ...REPLACEMENT PARTS 5 2 June 1997 Rosemount Analytical 748332 D NGA 2000 Non Dispersive Infrared Analyzer NOTES ...
Страница 58: ... blank ...
Страница 60: ... blank ...
Страница 62: ... blank ...