2.7.1 Configuring Nexus
To configure your messaging, each Nexus Station should be assigned a unique Station ID in the Settings
tab
–
this defaults to 01.
2.7.2 Transmitter Setup
The paging transmitter settings affect communication between the GP2009TR transmitter and GP2009N
pagers, or another GP2009TR paging repeater when used for remote displays.
It is recommended to set one frequency channel for each system on site to avoid signal traffic. Product
delivered from Rondish is centered on 450.375MHz, and should be kept within a band between
446.375MHz and 454.375MHz.
The configuration screen for GP2009TR is shown below with default settings:
TX Frequency: Pager system communication frequency (450.375 default)
RX Frequency: Pager system communication frequency (450.375 default)
Assign a 2-digit Nexus Station ID for each
Nexus terminal. The Station ID must be
unique for each station
Set Escalation Timer (in minutes) to
enable alarm escalation. If a call has not
been reset after this period of time it will
be escalated and made visible on other
displays or pagers. Escalation time ranges
from 1 to 99 min. Set the timer to 0 to
disable Call Escalation function.
Assign POCSAG codes for Local, Escalated,
Emergency and Code Blue events.