35
Calibrate Dens Menu
This menu is used to view and control the calibration of the X96S Density portion of the Gage. It contains
the following items:
ITEM
FUNCTION
State
Shows the state of the level configuration process
Low Ref Dens
Selecting this item takes the user to the
Reference
menu
High Cal Dens
Selecting this item takes the user to the
High Cal Dens
menu
Backgr.Counts
Shows and allows the user to set the
Backgr.Counts
. The background counts are
used with low detector count applications. This value used to calculate the indicated
level. Consult Factory before using this function. Typically used with
Low
Phase
Density Compensation.
Hi fault
Shows and allows the user to set the
Hi fault.
The Hi Fault is used for the density
compensation. When the density detector counts fall below this value, the
compensation is force to 1.0000
Lo fault
Shows and allows the user to set the
Lo fault.
The Hi Fault is used for the density
compensation. When the density detector counts goes above this value, the
compensation is force to 1.0000
Clear Ref/Cal
This item invokes a method that clears the level low reference and high calibration.
This allows the user to start over with out any stored reference or calibrated values.
State
is one of the following:
State
MEANING
Uncalibrated
Needs reference and calibrate.
Referenced
Needs calibrate
Partial Cal
Needs reference
Need Ref Density
Reference Density must be entered
Need Cal Density
Calibration Density must be entered
Fully Calibrated
Calibration complete
Invalid Data
Reference and calibrate data is inconsistent
Low Ref Density
is one of the following
State
MEANING
Reference
This item invokes a method that performs the low reference procedure
Ref Density
Shows and allows the user to set the reference density
Ref Cap
Shows the Referenced Captured Counts the last time the user performed a successful
low referenced
High Cal Density
is one of the following
State
MEANING
Calibrate
This item invokes a method that performs the high calibrate procedure
Cal Density
Shows and allows the user to set the reference density
Cal Cap
Shows the Calibrated Captured Counts the last time the user performed a successful
High Calibration.
1/uT
Shows the user the cal constant the X96S calculated during a High Calibration.