Low Ref Level
is one of the following
State
MEANING
Reference
This item invokes a method that performs the low reference procedure. After the
reference procedure is completed, it is important the user access the Ref Lev value and
insert or acknowledge the value.
Ref Level
Shows and allows the user to set the reference to the level of the process
Ref Cap
Shows the Referenced Captured Counts the last time the user performed a successful
low referenced
High Cal Level
is one of the following
State
MEANING
Calibrate
This item invokes a method that performs the high calibrate procedure. After the
calibrate procedure is completed, it is important the user access the Cal Lev value and
insert or acknowledge the value.
Cal Level
Shows and allows the user to set the calibration to the level of the process
Cal Cap
Shows the Calibrated Captured Counts the last time the user performed a successful
High Calibration.
Ref Mode
is one of the following:
Ref Mode
MEANING
Empty/Full
Vessel will be Empty (air) for reference and Full (filled with process) for calibration in
measuring area.
Process
Process material in measuring area (not necessarily empty and full for reference and
calibration). User will supply actual levels during reference and calibration.
Absorber
Absorber placed in radiation path.