Operation
•
Operate the lift using the pushbuttons on
the hand control. (figure 5)
•
For raising and lowering:
Directional arrows show the direction of
movement (up/down). The lift stops when
you release the buttons.
•
For base-width adjustment:
Press either of the two lower pushbuttons.
Emergency Stop
•
To use the Emergency Stop:
Push the red Emergency Stop button on
the control box. (figure 6)
•
To reset:
Turn the button in the direction of the
arrows.
Electrical emergency lowering
•
With a narrow object that is not sharp (a
key or allen wrench), press the hole
marked "EMERGENCY" on the control box.
(figure 6, arrow is pointing at holes
)
Mechanical emergency lowering
•
Pull the red emergency lowering regulator
upward one inch. (figure 7) The
mechanical emergency lowering device
functions only when the lift arm is loaded,
i.e., when the patient is in the lift.
Locking the wheels
•
To lock the wheels, press the brake lever
above the wheel down with your foot.
(figure 8)
•
During lifting, wheels should remain
unlocked
so that the lift may shift to the
patient's center of gravity. The wheels
should however be locked if there is a risk
for the lift moving into the patient, for
example when lifting from the floor.
Warning! Never move the lift by
pulling the actuator!
(figure 9)
5
6
7
8