RoMedic Rena 440 Скачать руководство пользователя страница 7

VER 3.0D 

07/2009 

 

 
 

 
Final Inspection 

 
 
 

 

 
 

 

 
 
 
 
 
 

 

 
 

 
 

 
 

• 

Final inspection of the lift is required 
before putting the lift into service. Once 

the lift has been removed from it’s 
shipping pallet, inspect it for signs of 

damage. 

• 

 Make sure all four casters are working 

properly, check castor locks (figure 1) 

• 

 Inspect all connections, nuts and bolts. 
Look at the packing material to see if any 

loose nuts, bolts, screws are discovered. If 

any are found identify where they go and 
put them back on.  

• 

Turn the emergency stop button clockwise 
so it pops out. (figure 2) 

• 

 Now the lift should be ready for 

operation. Take the hand control and push 
the up bottom. Run the lift arm all the way

up. Then push the down button and run 
the lift all the way down. 

• 

Verify the emergency stop button works 

by pushing it in and push the up or down 
button to verify the lift doesn’t work.  

• 

Follow the instructions for a sling that you 
are using and lift a person (that is not a 

patient)  in the lift to test operation.  

• 

Verify that emergency lowering works with 
someone in the lift. (figure 3) 

• 

If everything has checked out 
satisfactorily, plug the lift in for charging, 

verify the yellow charging light comes on. 

(figure 4) 

• 

View the video in-service DVD for proper 

training on the Rino before you put the lift 
into use with a patient.  

 

1

2

3

4

 

Содержание Rena 440

Страница 1: ...ers SystemRoMedic has four categories transfer repositioning support and lifting Visual inspection Inspect hoist functions regularly Check to ensure that material is free from damage Read instructions for all assistive devices used in connection with transfers Keep the manual where it is accessible to users of the product Do not leave the patient unattended during a lifting situation Before use Ma...

Страница 2: ...ce 12 Scale on DS model only 13 Note In this document the person being lifted is called the patient and the person helping is called the caregiver Warning This instruction guide contains information that is important for users of the product A complete understanding of the contents of the instruction guide is essential and only personnel who are well informed should use the equipment Remember to k...

Страница 3: ... positioned firmly and securely so that no part of the body can be injured the back belt is properly attached to the lift hook This should be checked when beginning to lift the patient when the sling straps are taut but the patient is still seated Warning Never leave the patient unsupervised in Rena Product Description 1 Handcontrol 2 Lift Arm 3 Handles 4 Mast 5 Battery Pack 6 Emergency Stop Butto...

Страница 4: ...ety nut Actuator base On EE DS models only 24 V 3 5 A permanent magnetic actuator Material Steel with hammer tone powder coat Emergency lowering Mechanical and electrical Wheels Front 75 mm dual Rear 100 mm with brake Intermittent Int Op 10 90 active operation max 2 operation min Out of a time of 100 active must be less than 10 though not more than 2 min Degree of protection IP 43 Built for indoor...

Страница 5: ...5 F 37 5 61 38 60 5 Row A B Row A B G 21 5 37 20 30 5 H 26 1 2 42 24 25 35 I 15 5 15 5 J 3 1 2 4 25 K 18 18 108 lbs 116lbs Note Row A is measurements in inches Row B is measurements in cm Weight of lift A lbs B kgs Note DS models weigh 15 lbs more A B C D F G H E I K RENA 440EM RENA 440EE DS RENA 440EM RENA 44OEE DS J ...

Страница 6: ...cm Row A B Row A B A 41 41 B 40 40 C 54 54 D 35 5 35 5 E 4 25 4 25 F 38 5 61 38 5 61 Row A B Row A B G 26 5 40 28 5 43 H 31 5 44 5 33 5 45 I 21 5 21 5 J 4 4 K 18 18 133 5 149 5 lbs A B C D F E Weight of lift A lbs B kgs Note DS models weigh 15 lbs more H I K G RENA 600EM RENA 600EE DS RENA 600EM RENA 600EE DS J ...

Страница 7: ...e lift should be ready for operation Take the hand control and push the up bottom Run the lift arm all the way up Then push the down button and run the lift all the way down Verify the emergency stop button works by pushing it in and push the up or down button to verify the lift doesn t work Follow the instructions for a sling that you are using and lift a person that is not a patient in the lift ...

Страница 8: ...key or allen wrench press the hole marked EMERGENCY on the control box figure 6 arrow is pointing at holes Mechanical emergency lowering Pull the red emergency lowering regulator upward one inch figure 7 The mechanical emergency lowering device functions only when the lift arm is loaded i e when the patient is in the lift Locking the wheels To lock the wheels press the brake lever above the wheel ...

Страница 9: ...third indication the battery is low is an audible beep as the lift is being operated If you hear a beeping sound the lift is below 25 charge level Charging methods Your Rena lift comes standard with a built in charger figure 12 Simply plug it into a electrical outlet to charge Verify the green LED light comes on to indicate the charger is receiving power Verify the yellow LED is on indicating char...

Страница 10: ...d that wall mounted chargers be plugged into surge protecting power strips StandUp Sling The RoMedic StandUp Sling Figure 16 comes in a variety of sizes from Small to XL The StandUp sling comes with a safety strap and available synthetic sheepskin padding Hygiene Sling The Hygiene sling figure 17 can be used with the Rena when the patient is unable or unwilling to bear weight With the removal of t...

Страница 11: ...dic distributor The legs of the lift won t operate or only go one way 1 Check that the cables are connected correctly and pushed in all the way The hand control cable can appear to be in but if it is not seated all the way one of the functions may not work If you are not sure pull it out and push it back in firmly 2 If lift still does not function correctly call your RoMedic distributor The lift s...

Страница 12: ...y authorized personnel IMPORTANT Repairs and maintenance may only be carried out by qualified service personnel using original RoMedic replacement parts or equivalent Transportation and storage During transportation or when the patient lift is not to be used for some time the emergency stop button should be pushed in Store the lift at a temperature above the freezing point and at normal relative h...

Страница 13: ... KG button it will read conv and hold until the display flashes Release the button figure 20 If the display flashes with a weight this indicates that a negative weight is being registered Display turns off after one minute of inactivity Press the recall button to display the last weight taken over 50lbs 22kg figure 20 Scale capacity is determined by weight capacity of lift Rena 440 is 440lbs Rena ...

Отзывы: