ROLA-CHEM

®

 Digital Controller                                        PN 554135      07/11/19                                       Page: 13 of 24 

4.3 Power panel access

  

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

4.3.1 Dip switches 

 

The controller has a series of dip switches to control functions that are not normally 
accessible on the front panel. These dip switches are located inside the power panel access 
cover. To access switches, remove power from the controller and remove power panel 
access cover. There are four dip switches- #1 is the top switch. The switch is ‘off’ when in the 
left position and ‘on’ in the right position. See appendix for settings. 
 

4.3.2 Fuses 

 
The ORP and pH are independently fused to protect the controller from defective feeders. 
These fuses are located inside the lower panel. The fuse holders are shipped with 5 amp 
fuses. (Replacement fuses are 5mm x 20mm Fast Acting IEC.) 
 

4.3.3  Flow Switch 

 
The controller has a flow switch terminal located inside the Power Panel Access Cover on the 
left side. It is shipped with a jumper in place. The terminal must either have the jumper or a 
flow switch installed. 
 
The flow switch light on the face of the controller is also a power light that shows that the 
controller is connected to a live power source.  If the terminal does not have a closed circuit, 
the controller will not turn on. 

15

18  EXTERNAL POW ER INPUT

13.  FLOW  SW ITCH TERMINAL
14.  DIP SW ITCH ARRAY
15.  FLOW  SW ITCH INPUT

12.  DATA PORT

17.  ORP OUTPUT

16.  pH OUTPUT

10.  TERMINAL STRIP
11.  pH FUSE

9.  ORP FUSE

17

18

16

12

14

13

FLOW

RS

4

ON

23

1

FUSE

pH

ORP

FUSE

9

9

11

10

Содержание RC554XP

Страница 1: ...f 24 F MODE FEED ONCE pH AUTO ORP LEVEL SET pH FEED ALERT CALIB pH FEED TIME HI ORP LOW ORP FEED L O W ALERT pH Operating Manual MODEL RC554XP ORP pH Digital Controller IMPORTANT SAFETY INSTRUCTIONS R...

Страница 2: ...lso remove power from the controller Provide a properly located outlet controlled by the circulating pump circuit 1 6 WARNING Risk of electric shock pinching chemical exposure Remove power sources or...

Страница 3: ...allation 4 1 Locating the controller 12 4 2 Mechanical installation 12 4 3 Power panel access 13 4 3 1 Dip switches 13 4 3 2 Fuses 13 4 3 3 Flow switch 13 4 3 4 Globe Flowcell 14 5 Operation 5 1 Start...

Страница 4: ...ings Press and release the MODE button until the desired ORP or pH LED is lit Once the settings are changed they will remain at those settings until they are changed again Set level press the SET LEVE...

Страница 5: ...isplay resolution The ORP set point is adjustable from 400 to 900 mV with default set point of 650 mV During an ORP feed cycle the ORP feed lamp will flash when sanitizer chemicals are being fed and i...

Страница 6: ...nternal terminal strip is provided for field wiring of the controller The internal step down transformer has a class two energy limiting rating to provide for electrical safety Note When automating an...

Страница 7: ...s Terminal strip for hardwire applications included Displays ORP and pH red digital Power and flow indicator LED Alert indicator Feed indicator LED ORP pH Mode indicator LED pH ORP Auto Note feed lamp...

Страница 8: ...onitors pH and ORP and displays programmed set points The Digital Display numbers will flash if Consecutive Feed Alert Limit is exceeded 5 pH STANDBY LAMP AND PROGRAMMING BUTTONS When pH standby lamp...

Страница 9: ...set desired pH parameters Note When scrolling through the set modes to reverse the direction simply release the button momentarily and press the button again Feed Once button By pressing the FEED ONCE...

Страница 10: ...calibration allows the adjustment of 1 pH in 1 pH increments Consecutive Feed Alert Limit The Digital Display numbers will flash and the pump feeder will not activate if the Set Level is not reached...

Страница 11: ...sed the ORP readout will scroll through the available ORP levels 100 mV to 640 mV in 5 mV increments and OFF Release the button when the desired level is reached The pH CALIB button is not functional...

Страница 12: ...oller in a location accessible to the public Install the controller in an area protected from natural elements sun rain freezing etc See appendix for typical feed system schematics 4 2 Mechanical Inst...

Страница 13: ...These fuses are located inside the lower panel The fuse holders are shipped with 5 amp fuses Replacement fuses are 5mm x 20mm Fast Acting IEC 4 3 3 Flow Switch The controller has a flow switch termina...

Страница 14: ...perate as a power light It is recommended the jumper be retained to assist in troubleshooting if necessary 4 3 4 Globe Flowcell see Globe Flowcell Operating Manual Connect the Globe Flowcell grounding...

Страница 15: ...description of controls Then select the desired Set Level Feed Time and High and Low Alert Levels Adjust Consecutive Feed Alert Limit if Set Level is not met within preset limit as indicated by the D...

Страница 16: ...troller with ROLA CHEM pumps the pump switch should be in the CONTINUOUS position not Timed or Off If the sanitizer level is too low or high adjust the ORP set level up or down in small increments for...

Страница 17: ...eed time is 0 6 sec on with 5 minute delay Feeders may require longer feed time If time is too short set points will not be maintained If feed time is too long overfeeding chemicals may occur Correct...

Страница 18: ...ight should be on F Press MODE to turn the controller On and return to operating mode The last two steps are extremely important because this puts the controller back in operating mode If these two st...

Страница 19: ...switch flow cell valves feeders tubing etc are not covered under this warranty Heavy duty probes are warranted from the date of original purchase for one 1 year ROLA CHEM under no circumstances shall...

Страница 20: ...in the timed feed mode 1 Turn off remove power then move 2 dip switch to ON 2 Re power press MODE to set it for ORP or pH 3 Note feed time setting then Set FEED TIME to OFF 4 Press FEED ONCE button t...

Страница 21: ...he buttons D The controller is now in test mode Test mode reduces the feed delay time from 5 minutes to 5 seconds to simplify diagnostics E Turn the controller Off again to exit test mode F Turn the c...

Страница 22: ...ROLA CHEM Digital Controller PN 554135 07 11 19 Page 22 of 24 8 4 Typical System Installations...

Страница 23: ...ROLA CHEM Digital Controller PN 554135 07 11 19 Page 23 of 24 8 4 Typical System Installations cont...

Страница 24: ...ROLA CHEM Digital Controller PN 554135 07 11 19 Page 24 of 24 NOTES...

Отзывы: