Command Reference
R&S
®
ZNB/ZNBT
873
User Manual 1173.9163.02 ─ 38
Manual operation:
See
"RFFE columns (sweep sequencer table)"
CONTrol<Ch>:SEQuence:CLEar:ALL
For unsegmented sweeps this command deletes the command/switch sequence for
the respective channel (defined using
CONTrol<Ch>:SEQuence<Nr>:RFFE<Bus>:
CONTrol<Ch>:SEQuence<Nr>:GPIO<Port>:VOLTage
Suffix:
<Ch>
.
Channel number
Usage:
Event
Manual operation:
See
CONTrol<Ch>:SEQuence:COUNt?
For unsegmented sweeps this command queries the length of the command sequence
defined for a particular channel (using
CONTrol<Ch>:SEQuence<Nr>:RFFE<Bus>:
).
Suffix:
<Ch>
.
Channel number
Usage:
Query only
Manual operation:
See
CONTrol<Ch>:SEQuence<Nr>:DELay
<Delay>
For unsegmented sweeps this command allows to introduce delays between subse-
quent sequence steps and between the final sequence step and the sweep (segment)
start.
The command/switch sequences are defined using
CONTrol<Ch>:SEQuence<Nr>:GPIO<Port>:
.
Suffix:
<Ch>
.
Channel number
<Nr>
Sequence number, defining the order in which the commands
shall be executed.
For every channel and RFFE interface, the sequence numbers
must be consecutive, starting at 1; reusing a sequence number
overwrites a previously defined command.
Parameters:
<Delay>
Delay time
Default unit: s
Manual operation:
See
"Wait (Sweep Sequencer Table)"
SCPI Command Reference