Command Reference
R&S
®
ZNB/ZNBT
1194
User Manual 1173.9163.02 ─ 38
Example:
INP2:ATT 10
Set the step attenuator for the wave received at port 2 and for all
channels to 10 dB. The waves at the other test ports are not
affected.
SENS1:POW:ATT? BREC
Query the receiver step attenuator setting at port 2 and for chan-
nel no. 1. The response is
10
.
INSTrument[:SELect]
<Channel>
Selects a channel
<Ch>
as active channel. To select a channel number > 4 use the
generalized command
Parameters:
<Channel>
CHANnel1 | CHANnel2 | CHANnel3 | CHANnel4
Number of the channel to be activated. The channel must be
created before using
ON
.
Range:
CHANNEL1, CHANNEL2, CHANNEL3, CHAN-
NEL4
*RST:
CHANNEL1
Example:
CONF:CHAN2:STAT ON; :INST CHANNEL2
Create channel no. 2 and select it as the active channel.
OUTPut<Chn>:DPORt
<Port>
Selects a source port for the stimulus signal (drive port). The setting acts on the active
trace. The effect of the drive port selection depends on the measurement parameter
associated to the active trace:
●
If an S-parameter S
<out><in>
is measured, the second port number index <in> (input
port of the DUT = drive port of the analyzer) is set equal to the selected drive port:
Drive port selection affects the measured quantity.
●
If a wave quantity or a ratio is measured, the drive port is independent from the
measured quantity:
Note:
to address
port numbers > 4.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Port>
PORT<no>
Physical port number.
In "set direction" only constants PORT1 to PORT4 can be used.
*RST:
PORT1
R&S ZVR/ZVAB Compatible Commands