Command Reference
R&S
®
ZNB/ZNBT
1190
User Manual 1173.9163.02 ─ 38
Suffix:
<Chn>
.
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10.
Parameters:
<SearchValue>
Target search value of marker no.
<Mk>
.
Range:
Depending on the format of the active trace. For a
dB Mag trace the range is -200 dB to +200 dB.
Increment: 0.1 dB
*RST:
Depending on the trace format; 0 dB for a dB Mag
trace.
Default unit: dB
CALCulate<Chn>:MARKer<Mk>:FUNCtion[:SELect]
<Mode>
Selects a search mode for marker no.
<Mk>
, which can then be initiated using one of
the
CALCulate<Ch>:MARKer<Mk>:SEARch...
,
CALCulate<Ch>:MARKer<Mk>:MAXimum
or
CALCulate<Ch>:MARKer<Mk>:MINimum
functions. The marker must be created
before using
CALCulate<Chn>:MARKer<Mk>[:STATe]
.
Note:
This command is not needed except for compatibility with ZVR programs. Use
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
to select a search mode and
at the same time initiate the search. The
CALCulate<Ch>:MARKer<Mk>:SEARch...
,
CALCulate<Ch>:MARKer<Mk>:MAXimum
or
CALCulate<Ch>:MARKer<Mk>:MINimum
functions also select the search mode.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10. For a bandfilter search
(
BFILter
) this numeric suffix is ignored and may be set to any
value because the bandfilter search functions always use mark-
ers M1 to M4.
Parameters:
<Mode>
MAXimum | MINimum | RPEak | LPEak | NPEak | TARGet |
LTARget | RTARget | BFILter | MMAXimum | MMINimum |
SPRogress
See
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
CALCulate<Chn>:MARKer<Mk>:MAXimum
CALCulate<Chn>:MARKer<Mk>:MINimum
These commands select a search mode for marker no.
<Mk>
and initiate a maximum
and minimum search, respectively. The marker must be created before using
CALCulate<Chn>:MARKer<Mk>[:STATe]
ON
.