Command Reference
R&S
®
ZNB/ZNBT
1166
User Manual 1173.9163.02 ─ 38
Example:
SYST:PRES:SCOP SING
Define the scope of a preset: the active recall set is reset only.
SYST:FPR
Reset the parameters of the current recall set and delete chan-
nel 1.
Usage:
Event
SYSTem:FREQuency?
<MinMax>
Queries the minimum and maximum frequency of the network analyzer. For an over-
view refer to the tables at the beginning of
Chapter 7.3.15.10, "[SENSe:]FRE-
Tip:
In contrast to
and the other sweep range
commands,
SYSTem:FREQuency?
can be used in all sweep modes.
Query parameters:
<MinMax>
MINimum | MAXimum
Return values:
<Frequency>
MINimum | MAXimum
Return minimum or maximum frequency.
Example:
See
Usage:
Query only
Manual operation:
See
"Start Frequency / Stop Frequency / Center Frequency /
SYSTem:IDENtify[:STRing]
<InstId>
Defines an identity string for the network analyzer. The query is equivalent to
*IDN?
.
Parameters:
<InstId>
String parameter containing the instrument identity
Example:
See
Manual operation:
See
SYSTem:IDENtify:FACTory
Resets the response to the
*IDN?
query to the factory default value. This command
overwrites a user-defined identification string; see example.
Example:
SYSTem:IDENtify:STRing 'MyDevice'; *IDN?
Define an identity string. The response is
MyDevice
.
SYSTem:IDENtify:FACTory; *IDN?
Re-activate the factory setting. The response is
Rohde-Schwarz,<instrument type>-<ports>,<stock
no.><serial no.>,<firmware version>
.
SCPI Command Reference