Command Reference
R&S
®
ZNB/ZNBT
1029
User Manual 1173.9163.02 ─ 38
Example:
COUP NONE
Activate the alternated sweep mode.
TRIG:LINK 'PPO'
Set the triggered measurement sequence equal to one partial
measurement. Each trigger event starts one partial measure-
ment for all sweep points.
Manual operation:
See
7.3.15.9
[SENSe:]DC...
The
[SENSe:]DC...
commands select and configure the DC voltage measurement.
[SENSe<Ch>:]DC<DCInp>:RANGe
<Voltage>
Assigns a DC voltage range to the rear panel input connector DC INPUT <DCInp>.
Suffix:
<Ch>
.
Channel number
<DCInp>
Number of DC INPUT connector on the rear panel, 1 to 4
Parameters:
<Voltage>
DC voltage range. The entered numeric value is rounded up to
±
300 mV,
±
3 V,
±
20 V. Values above +20 V are rounded down.
Default unit: V
Example:
SENSe1:DC2:RANGe 1; :DC2:RANGE?
Select a DC measurement at DC INPUT2 with an expected DC
voltage range of
±
1 V. The value is rounded up; the response is
3
.
Manual operation:
See
7.3.15.10
[SENSe:]FREQuency...
The
[SENSe:]...
commands set frequency-related parameters, especially the mea-
surement and display ranges for the different sweep types. The frequency ranges for
the different instrument models are listed below; for more details refer to the data
sheet.
Table 7-13: Frequency ranges of R&S
ZNB/ZNBT analyzers
Frequency set-
tings
R&S
ZNB4
R&S
ZNB8 /
R&S
ZNBT8
R&S
ZNB20 /
R&S
ZNBT20
R&S
ZNB40
variant 72
R&S
ZNB40
variants 82
and 84
Start, Stop
9 kHz to 4.5
GHz
9 kHz to 8.5
GHz
100 kHz to 20
GHz
10 MHz to 40
GHz
100 kHz to 40
GHz
Center
>9 kHz to <4.5
GHz
>9 kHz to <8.5
GHz
>100 kHz to
<20 GHz
>10 MHz to <40
GHz
>100 kHz to
<40 GHz
Span
1 Hz to
4.499991 GHz
1 Hz to
8.499991 GHz
1 Hz to 19.9999
GHz
1 Hz to 39.99
GHz
1 Hz to 39.9999
GHz
SCPI Command Reference