Command Reference
R&S
®
ZNB/ZNBT
1089
User Manual 1173.9163.02 ─ 38
Parameters:
<PortString>
String representation '<Source>:<a wave>:<b wave>' where
<Source> =
i
refers to (original) VNA ports
<Source> = G
i
refers to external Generator ports
<a wave>, <b wave> is one of the original a
i
or b
i
,
without reusing the same port in different (redefined) ports
Example:
SENSe:UDSParams:ACTive OFF
resolves all redefined ports
SENSe:UDSParams1:PARam '1:b2:b3'
redefines physical port 1 as source = (original) physical port 1,
reference receiver = physical port 2, measurement receiver =
physical port 3
SENSe:UDSParams:ACTive?
)
Manual operation:
See
7.3.16 SOURce Commands
The
SOURce...
commands affect the source settings of the R&S
ZNB/ZNBT.
7.3.16.1
SOURce:FREQuency...
The
SOURce:FREQuency...
commands configure the sources for frequency conver-
sion measurements and control the frequency and power of the internal signal source.
SOURce<Ch>:FREQuency:CONVersion:MIXer:PMFixed
.................................................1089
SOURce<Ch>:FREQuency:CONVersion:MIXer:PMODe
.................................................. 1090
SOURce<Ch>:FREQuency<PhyPt>[:CW]
...................................................................... 1090
SOURce<Ch>:FREQuency<PhyPt>:FIXed
..................................................................... 1090
SOURce<Ch>:FREQuency<PhyPt>:CONVersion:ARBitrary:EFRequency<Gen>
SOURce<Ch>:FREQuency<PhyPt>:CONVersion:ARBitrary:IFRequency
.......................... 1093
SOURce<Ch>:FREQuency:CONVersion:MIXer:PMFixed
<arg0>[, <arg1>]
Assigns a fixed power to the RF, LO 1, LO 2, or to the IF signal.
Suffix:
<Ch>
.
Channel number
Parameters:
<arg0>
RF | LO | LO1 | LO2 | IF
RF – mixer input signal
LO | LO1 – local oscillator signal no. 1
LO2 – local oscillator signal no. 2, for 2-stage mixer measure-
ments (
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:
IF – mixer output signal
SCPI Command Reference