Command Reference
R&S
®
ZNB/ZNBT
1091
User Manual 1173.9163.02 ─ 38
The frequency range depends on the R&S
Note:
SOURce<Ch>:FREQuency<PhyPt>[:CW]|:FIXed
is equivalent to
[SENSe<Ch>:]FREQuency[:CW]|:FIXed
. Source and receiver frequency are
always equal; the four commands overwrite each other.
With option R&S
ZNB/ZNBT-K4, Frequency Conversion Measurements, port-specific
source and receiver frequencies can be defined; see
FREQuency<PhyPt>:CONVersion:ARBitrary:IFRequency
.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Test port number of the analyzer. This suffix is ignored because
the selected frequency applies to all source ports in the active
channel.
Parameters:
<FixedFreq>
Fixed stimulus and analyzer frequency.
*RST:
1 GHz
Default unit: Hz
Example:
FUNC "XPOW:POW:A1"
Activate a power sweep and select the wave quantity a
1
as mea-
sured parameter for channel and trace no. 1.
SOUR:FREQ:CW 100 MHz
Set the CW frequency to 100 MHz.
Manual operation:
See
SOURce<Ch>:FREQuency<PhyPt>:CONVersion:ARBitrary:EFRequency<Gen>
<Boolean>, <Numerator>, <Denominator>, <Offset>, <SweepType>
Defines an external generator frequency for frequency-converting measurements. The
external generator frequency is either a range (for frequency sweeps) or a CW fre-
quency (for power, time and CW Mode sweeps).
Suffix:
<Ch>
.
Channel number
<PhyPt>
Test port number of the analyzer. This suffix is ignored because
the generator is referenced via
<Gen>
.
<Gen>
Generator number
Parameters:
<Boolean>
Switch the generator on or off.
SCPI Command Reference