Command Reference
R&S
®
ZNB/ZNBT
1073
User Manual 1173.9163.02 ─ 38
[SENSe<Ch>:]SEGMent<Seg>:POWer:GAINcontrol:CONTrol
<Boolean>
Defines whether common or independent gain control settings shall be used for the
individual sweep segments.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number. This suffix is ignored.
Parameters:
<Boolean>
ON
–
use independent AGC settings, to be defined via
[SENSe<Ch>:]SEGMent<Seg>:POWer:GAINcontrol
OFF
–
reset the AGC in all sweep segments to the AGC for
unsegmented sweeps, defined via
[SENSe<Ch>:]POWer:GAINcontrol:
. ON will not restore the previous values.
*RST:
ON
Example:
See
[SENSe<Ch>:]SEGMent<Seg>:POWer:GAINcontrol
Manual operation:
See
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel]:CONTrol
<Boolean>
Selects common or independent internal source "Power" settings for the sweep seg-
ments.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number. This suffix is ignored; the setting con-
trols the whole segmented sweep.
Parameters:
<Boolean>
ON
–
use independent power settings, to be defined via
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel]
OFF
–
reset the power in all sweep segments to the power for
unsegmented sweeps, defined via
POWer<PhyPt>[:LEVel][:IMMediate][:AMPLitude]
will not restore the previous values.
The parameter is automatically switched to
ON
when a segment
power is entered using
*RST:
OFF
SCPI Command Reference